Skip to main content
. 2019 May 28;9:7938. doi: 10.1038/s41598-019-44128-z

Table 2.

Recoveries and detection limits for amorphous SiO2-NPs.

Detection
method
Digestion
method
SiO2
source
Matrix Recovery ± s.d.
(%)
BEC LOQ LOD
(µg SiO2 L−1)
ICP-OES KOH0.1 colloidal water 85 ± 2 24.5 80.1 41.1
KOH0.5 DMEM 84 ± 20 48.2 216.3 98.6
KOH1.0 fumed water 114 ± 25 63.7 67.6 64.8
colloidal tomato saucea 124 ± 5 n.d. n.d. n.d.
food grade seasoninga 95 ± 13 n.d. n.d. n.d.
RT + KOH1.0 colloidal water 84 ± 4.9 113 296 168
fumed 105 ± 1.4 113 296 168
Colorimetry RT + KOH0.1 colloidal 111 ± 7 115 264 148
fumed 67 ± 2 22 430 144
RT + KOH1.0 76 ± 9 120 307 176

Colloidal Stöber SiO2-NPs; fumed Aerosil® SiO2-NPs; and food grade SiO2, also known as E551. The SiO2-NPs were analyzed in matrices of increasing complexity: water, cell culture medium (DMEM), potato seasoning, and tomato sauce. The numbers in the digestion method names are the KOH concentrations used in mol L−1.

s.d.: standard deviation. n.d.: not determined. BEC: background equivalent concentration. LOD: instrument limit of detection. LOQ: instrument limit of quantification. RT: room temperature. ICP: Inductively coupled plasma – optical emission spectrometry. aThe food matrix was pre-digested as mentioned in section Pre-digestion of samples containing food matrix. Potato seasoning and tomato sauce recovery relative to mass after pre-digestion of the food matrix.