Table 5.
Coefficients of variations of the ventilatory variables measured with the non-contact vibrometer using airborne ultrasound during synchronous pneumotachograph-SONAR recordings (SONAR-PNT) and with the SONAR only (SONAR-only).
| SONAR signal | TT S | TI S | TE S | Displacement cm |
|---|---|---|---|---|
| In SONAR-PNT condition | 0.09 (0.08–0.12) | 0.12 (0.10–0.16) | 0.11 (0.10–0.16) | 0.18 (0.15–0.22) |
| In SONAR-only condition | 0.11 (0.08–0.15) | 0.15 (0.14–0.16) | 0.14 (0.11–0.22) | 0.23 (0.16–0.29) |
| p-Value | 0.74 | 0.18 | 0.52 | 0.19 |
Data are presented as median (Q1–Q3). TT, breathing cycle period; TI, inspiratory time; TE, expiratory time; s, second.