Performance Measures between Icobrain and Expert Reference Values for Acute Intracranial Lesions and Cistern Volumes and for Midline Shift

  Acute intracranial lesions Cisterns
Volume difference (mL) 0.07 (−7.73, −7.47) –0.01 (−1.66, −1.21)
Absolute volume difference (mL) 8.83 (2.02–17.05) 1.48 (0.70–2.26)
Classification accuracy 0.92
ICC 0.91 0.94
Dice 0.73 (0.55–0.81) 0.70* (0.66–0.73)
Precision 0.75 (0.63–0.87) 0.72* (0.67–0.79)
Sensitivity 0.75 (0.61–0.84) 0.69* (0.64–0.74)
  Midline shift  
Shift difference (mm) –0.22 (−0.88, −0.76)  
Absolute shift difference (mm) 0.86 (0.37–1.48)  
Classification accuracy 0.89  
ICC 0.93  
*

For cisterns, Dice, sensitivity, and precision are reported for volume >5 mL.

Except classification accuracy and ICC, all other measures are presented in median (25th-75th) percentiles.