Published in final edited form as: Anal Chem. 2019 Mar 13;91(7):4656–4664. doi: 10.1021/acs.analchem.9b00021

Table 1.

Variables Tested for TDI-SFC to Improve TDI Synchronization and Fluorescence Sensitivity

TDI-SFC
variable
range improvement figure
cell focusing none → 1D focusing synchronization:4.5 × Speak,0.8× Δt 3
cell velocity 1.41 → 1.83 ← 2.11 mm/s synchronization: 2.3 × Speak, 0.5× Δt 4
excitation power 4 → 21 mW sensitivity: >9× Sint 5A
objective NA 0.5 → 0.75 NA sensitivity: 8–10× Sint, 4× SNR 5B,C
TDI-integration time 200 → 400 ms sensitivity: 2× Sint, 1 × SNR 5D