Table 2.

Electrode characteristics and ANSI/AAMI requirements for disposable ECG electrodes

Textile ANSI/AAMI
DC resistance < 0.1 KΩ < 2 KΩ
AC impedance (12) 1.26 ± 0.18 KΩ < 3 KΩ
Internal noise 1–3 μV < 150 μV
Defibrillation discharge at 200 V
 Recovery PV 0.195 ± 0.80 mV < 100 mV
 Rate of change of PV No change < 1 mV/sec
 AC impedance after test 1.14 ± 0.07 KΩ < 3 KΩ
 DC offset voltage (12) 0.0028 ± 0.0020 mV < 100 mV

Data are reported as the means ± SD

ANSI American National Standard Institute, AAMI Association for the Advancement of Medical Instrumentation, ANSI/AAMI standards: Ref. [14], PV depolarizing voltage, n: number of experiments