Table 2.

Precision and accuracy for the determination of BAC in bulk form using the proposed spectrophotometric methods.

Method Nominal value (μg/mL) Found ± SDa (μg/mL) RSD (%)b Er (%)c
Method I Within-day 15 14.71 ± 0.23 1.56 –1.93
25 25.34 ± 0.15 0.59 01.36
35 35.21 ± 0.32 0.91 00.60
Between-day 15 14.96 ± 0.30 2.01 –0.27
25 25.08 ± 0.07 0.28 00.32
35 34.97 ± 0.60 1.72 –0.09
Method II Within-day 10 10.05 ± 0.09 0.90 00.50
15 15.16 ± 0.17 1.12 01.07
20 20.16 ± 0.33 1.64 00.80
Between-day 10 10.07 ± 0.15 1.49 00.70
15 14.96 ± 0.28 1.87 –0.27
20 20.34 ± 0.38 1.87 01.70
Method III Within-day 10 09.92 ± 0.03 0.30 –0.80
15 15.20 ± 0.15 0.99 01.33
20 19.59 ± 0.16 0.82 –2.05
Between-day 10 9.97 ± 0.02 0.20 –0.30
15 14.91 ± 0.17 1.14 –0.60
20 19.82 ± 0.04 0.20 –0.90
a

Mean ± standard deviation for three determinations.

b

% Relative standard deviation.

c

% Relative error.