Table 2.

Calibration parameters, sensitivity, recovery, accuracy and precision of the optimized LC-QqQ/MS method.

Compound Calibration range LLOD (ng/ mL) LLOQ (ng/ mL) Linear dynamic range R2 Spiked amount (ng/ mL) Accuracy (%) Precision (%) Recovery (%)
GA Low 0.35 1.02 5.100–2040 0.998 200 −3.82 4.38 115 (8.82)
High 2040–6120 0.995 2000 −17.1 6.53 121 (10.2)
3,4-diHBA Low 0.62 2.55 2.560–2040 0.998 200 3.31 3.21 92.0 (7.41)
High 2040–6120 0.999 2000 7.49 3.65 102 (5.32)
3,4-diHPAA Low 1.23 5.01 2.652–2080 0.999 200 −15.5 8.96 108 (12.0)
High 2080–6240 0.996 2000 −17.7 4.13 116 (6.34)
4-HBA Low 0.29 0.97 1.325–1060 0.999 200
High 1060–6360 0.998 2000 −9.23 4.55 102 (1.87)
( + )-c Low 0.41 1.32 1.325–1060 0.999 200 4.57 9.69 91.2 (2.78)
High 1060–6360 0.996 2000 2.89 4.54 95.0 (6.44)
3,4-diHPPA Low 0.51 2.60 5.204–1040 0.998 200 −8.75 12.2 110 (9.43)
High 1040–6240 0.997 2000 −20.8 3.16 117 (6.56)
HA Low 0.42 1.29 2.505–1002 0.999 200
High 1002–6012 0.998 2000 −3.57 3.77 114 (9.02)
VA Low 0.75 2.36 2.602–1040 0.999 200 5.82 0.98 104 (2.84)
High 1040–6240 0.999 2000 2.21 4.05 102 (8.04)
CA Low 0.25 1.08 2.464–980.1 0.999 200 −11.0 9.55 102 (2.90)
High 980.1–6020 0.997 2000 −7.81 4.44 108 (4.15)
( − )-EC Low 0.51 1.60 1.275–1020 0.999 200 12.6 5.81 93.6 (4.59)
High 1020–6120 0.998 2000 14.3 5.35 91.1 (7.18)
HVA Low 0.63 1.62 2.700–2160 0.999 200 −3.28 5.90 100 (5.83)
High 2160–6480 0.999 2000 −6.48 1.58 106 (2.62)
3-HBA Low 1.25 3.75 2.452–1020 0.999 200 −0.67 5.12 108 (5.42)
High 1020–6120 0.998 2000 −1.96 2.81 102 (2.25)
3-HPAA Low 0.74 2.07 2.718–1040 0.999 200 −3.10 4.33 106 (6.78)
High 1040–6240 0.999 2000 −1.85 1.76 94.6 (4.10)
diHCA Low 2.05 8.11 10.80–1080 0.997 200 −15.9 4.70 112 (5.45)
High 1080–6480 0.995 2000 −19.9 2.19 116 (4.73)
p-CA Low 0.22 0.69 1.245–996 0.999 200 −1.24 1.89 101 (5.14)
High 996–5976 0.996 2000 −5.74 4.22 100 (3.45)
FA Low 0.50 1.51 1.374–1080 0.999 200 4.96 4.67 96.6 (5.80)
High 1080–6480 0.999 2000 6.04 7.84 93.6 (5.15)
3-HPPA Low 0.31 2.65 5.306–1060 0.999 200
High 1060–5360 0.998 2000 4.27 3.19 90.2 (4.94)
4-HPVA Low 0.76 2.61 2.607–1040 0.999 200 6.20 3.40 94.4 (4.13)
High 1040–5240 0.996 2000 8.25 2.93 96.8 (2.20)
trans-Cinnamic acid-d7 (I.S.) 0.23 0.71 1.272–1018 0.999

Data are expressed as means of six independent replicates. Standard deviations for Recoveries (%) were shown in parentheses.

Abbreviations are the same as in Table 1. LLOD, lower limit of detection; LLOQ, lower limit of quantification.

Calibration concentrations were divided into two ranges (low, ca. 1.25–1000 ng/mL and high, ca. 1000–6000 ng/mL), with optimum fit-to-line linear response. Some values for the 200 ng/mL spiked samples were null (−) due to the high response from the amount of the compounds existing endogenously.