Table 2. Carryover testing results on Seal AA3 HR analyser.

Urine sample 1st day
UI, µg/L
2nd day
UI, µg/L
3rd day
UI, µg/L
1st cup low, 1st time 38.1 36.5 37.8
1st cup low, 2nd time 39.2 38.2 38.1
1st cup low, 3rd time 35.6 37.4 37.6
1st cup high, 1st time 503.6 490.3 492.1
1st cup high, 2nd time 485.7 492.4 489.5
1st cup high, 3rd time 490.6 495.4 492.4
2nd cup low, 1st time 38.4 37.6 38.4
2nd cup low, 2nd time 37.6 36.8 36.1
2nd cup low, 3rd time 38.2 37.5 38.9
UI - Urinary iodine. low - sample with low urinary iodine concentrations. high - sample with high urinary iodine concentrations.