Table 2. Carryover testing results on Seal AA3 HR analyser.
| Urine sample |
1st day UI, µg/L |
2nd day UI, µg/L |
3rd day UI, µg/L |
|---|---|---|---|
| 1st cup low, 1st time | 38.1 | 36.5 | 37.8 |
| 1st cup low, 2nd time | 39.2 | 38.2 | 38.1 |
| 1st cup low, 3rd time | 35.6 | 37.4 | 37.6 |
| 1st cup high, 1st time | 503.6 | 490.3 | 492.1 |
| 1st cup high, 2nd time | 485.7 | 492.4 | 489.5 |
| 1st cup high, 3rd time | 490.6 | 495.4 | 492.4 |
| 2nd cup low, 1st time | 38.4 | 37.6 | 38.4 |
| 2nd cup low, 2nd time | 37.6 | 36.8 | 36.1 |
| 2nd cup low, 3rd time | 38.2 | 37.5 | 38.9 |
| UI - Urinary iodine. low - sample with low urinary iodine concentrations. high - sample with high urinary iodine concentrations. | |||