Table 2.

Left column: main parameters relevant for the operating regime of the X-ray source; the electron bunch and scattering laser pulse figures were used as input of the Thomson scattering simulation code TSST.

Thomson scattering X-ray source parameters Imaging simulation parameters
Electron bunch SAD 1000 mm
Size 2 μm × 2 μm (tr), 3 μm (long) ADD 1500 mm
Divergence 10 mrad Magnification 2.5
Bunch charge 100 pC Detector size (transverse × axial) 50 × 25 mm2
Mean energy ~50MeV Detector pixel size 250 μm
Energy spread 20% rms X-ray pulse duration 0
Scattering laser pulse Total n. of X-ray pulses (for the entire 4D reconstruction) 6000 to 24000
Duration 500 fs (chirped) Unattenuated no. photons per 250 μm pixel 5000
Energy 1J Unattenuated no. photons per 50 μm pixel 200
Central wavelength 800 nm Size of the cubic voxel in tomographic reconstructions 100 μm
Spot size 20 μm
a 0 ~0.24
X-ray source
Duration ~10fs
Source transverse size 2 μm × 2 μm
No. of photons within 10 mrad per laser shot ~9.8×107

More parameters related to a possible LWFA regime able to provide the required electron bunch are given in the Supplementary Materials. Right column: parameters of the numerical simulation with the 4D mouse chest phantom (SAD: Source to axis distance; ADD: Axis to detector distance).