. 2019 May 16;20(6):91–98. doi: 10.1002/acm2.12611

Table 1.

Recommendations for intrafraction motion monitoring during stereotactic radiotherapy with optical surface monitoring system

Issue Recommendations
Thermal drift Determine time for thermal drift equilibrium
Turn on projectors before patient treatment
Couch walk‐out Measure couch walk‐out (EPID, dial gauge, etc.)
Phantom study to compare walk‐out with real time deltas (RTDs)
RTD dependence on surface‐to‐isocenter distance Advanced camera optimization upgrade
Increase in RTDs at non‐zero couch angles Perform cold calibration detailed in customer information bulletin
RTD exceeds thresholds at non‐zero couch angles Return couch to zero and evaluate RTDs
RTD exceeds thresholds at couch zero Repeat radiographic image guidance