Slit camera measurement: small focal spot sizes (nominal focal spot size: 0.60 mm) measured at various exposure levels, compared between those of the screen‐film (SF) and computed radiography (CR) systems.

Tube voltage (kVp) mAs Exposure (mR) Width (mm) d (%)a Length (mm) d (%)a
SF CR SF CR
60 2.4 8.30 0.80 0.83 3.75 1.20 1.18 −1.67
60 3.0 12.27 0.80 0.78 −2.50 1.20 1.18 −1.67
70 1.8 6.17 0.80 0.78 −2.50 1.20 1.16 −3.33
90 1.2 7.27 0.80 0.83 3.75 1.20 1.23 2.50
CV (%)     0.0% 3.1%   0.0% 2.2%  
a