. 2019 May 21;16:e00124. doi: 10.1016/j.plabm.2019.e00124

Table 2.

Diagnostic accuracy of urinary reagent strip test when compared with reference standard values of parameters.

CSF LEUCOCYTE (95% CI) CSF PROTEIN (95% CI) CSF GLUCOSE (95% CI) CSF ERYTHROCYTE (95% CI)
True Positive 117 285 315 99
False Positive 09 33 03 00
False Negative 00 03 06 00
True Negative 234 39 36 261
Sensitivity 100% (90.9%–100%) 98.9% (94.3%–99.9%) 98.1% (93.4%–99.8%) 100% (89.4%–100%)
Specificity 96.3% (89.6%–99.2%) 54.2% (32.8%–74.5%) 92.3% (63.9%–99.8%) 100% (95.9%–100%)
Positive Likelihood Ratio 27 (8.9–81.9) 2.2 (1.4–3.34) 12.8 (1.9–83.8) 00
Negative Likelihood Ratio 00 0.02 (0.00–0.14) 0.02 (0.01–0.08) 00
Disease prevalence 32.5% (24.2%–41.6%) 80% (71.72%–86.75%) 89.2% (82.2%–94.1%) 27.5% (19.8 %–36.4%)
Positive Predictive Value (PPV) 92.9% (80.5%–98.5%) 89.6% (82.19%–94.70%) 99.1% (94.9%–99.9%) 100% (89.4%–100%)
Negative Predictive Value (NPV) 100% (95.38%–100.00%) 92.9% (66.13%–99.82%) 85.7% (57.2%–98.2%) 100% (95.9%–100%)
Area under the ROC curve (AUC) 0.92 0.714 0.887 1.000
Standard Error 0.028 0.0705 0.0828 0.000