. 2019 Mar 15;11(5):333–347. doi: 10.4155/bio-2018-0202

Table 4. . Accuracy and precision for chloroquine and desethylchloroquine extracted from human EDTA whole blood, plasma and dried blood sample.

Sample type Chloroquine Desethylchloroquine

  Nominal concentration (ng/ml) Measured concentration (ng/ml) Accuracy (%) Precision (%) Nominal concentration (ng/ml) Measured concentration (ng/ml) Accuracy (%) Precision (%)

        Between run CV (interassay) Within run CV (intra-assay)       Between run CV (interassay) Within run CV (intra-assay)
Whole blood (n = 5, b = 4)

LLOQ 2.56 2.63 103 7.27 5.13 3.36 3.25 96.9 6.97 4.14

QC1 7.56 7.70 102 3.46 2.74 9.40 9.70 103 3.92 2.57

QC2 91.5 94.9 104 2.11 2.78 102 106 104 2.79 1.60

QC3 1049 1093 104 2.49 1.92 1049 1090 110 2.78 1.58

ULOQ 1220 1233 101 3.40 3.59 1220 1244 102 2.49 3.76

Over curve 2556 2681 104 2.52 2.77 2556 2650 104 5.68 2.62

DBS (n = 5, b = 6)

LLOQ 1.82 1.95 107 16.8 14.0 2.95 3.21 109 15.7 12.4

QC1 6.03 5.76 95.5 5.20 7.09 8.89 8.40 94.5 14.2 7.79

QC2 102 97.9 96.0 5.90 4.29 124 112 90.2 7.97 4.48

QC3 1334 1323 99.2 4.56 5.21 1334 1285 96.4 7.65 3.62

ULOQ 1552 1590 102 4.29 4.54 1552 1566 101 3.11 3.26

Over curve 3429 3511 102 6.97 3.41 3429 3453 101 6.30 5.25

Plasma (n = 5, b = 4)

LLOQ 1.41 1.45 103 10.1 9.64 1.41 1.43 101 6.96 12.6

QC1 4.64 4.35 93.7 5.65 6.14 4.64 4.34 93.5 3.40 5.21

QC2 56.4 55.3 98.1 5.94 3.71 56.4 53.0 94.0 3.88 3.98

QC3 524 509 97.2 6.51 3.28 524 493 94.1 4.78 4.22

ULOQ 610 597 97.9 7.53 4.02 610 598 98.0 1.31 3.51

Over curve 1932 1953 101 5.00 2.87 1932 1939 100 5.12 4.13

Over curve, that is, sample dilution integrity test.

Over-curve whole blood sample diluted five-times with blank plasma.

b: Number of runs; DBS: Dried blood spot; n: Number of samples in each run; QC: Quality control.