Table III.
Parameters of liquid chromatography-tandem mass spectrometry method for measuring glutamate, aspartate, glutamine and asparagine.
Parameter | Value |
---|---|
Mobile phase | Water containing 0.05% (v/v) formic acid |
Column | Phenomenex Kinetex F5 column (4.6×100 mm, 2.6 µm) |
Column temperature (°C) | 30 |
Flow rate (ml/min) | 0.3 |
Capillary voltage (kV) | 3.5 |
Drying gas temperature (°C) | 350 |
Drying gas flow (l/min) | 10 |
Nebulizer pressure (psi) | 40 |
MRM transition (m/z) | |
Glu | 148→84 |
Gln | 147→83 |
Asp | 134→88 |
Asn | 133→87 |
Glu-IS | 151→87 |
Asp-IS | 137→91 |
Dwell (msec) | |
Glu | 100 |
Gln | 100 |
Asp | 100 |
Asn | 100 |
Glu-IS | 100 |
Asp-IS | 100 |
Fragmentor (V) | |
Glu | 80 |
Gln | 80 |
Asp | 50 |
Asn | 50 |
Glu-IS | 80 |
Asp-IS | 50 |
CE (eV) | |
Glu | 18 |
Gln | 5 |
Asp | 7 |
Asn | 4 |
Glu-IS | 18 |
Asp-IS | 7 |
Glu-IS, 2H3-Glutamate; Asp-IS, 2H3-Aspartate; CE, Collision Energy