Skip to main content
. 2019 Apr 30;18(1):188–198. doi: 10.3892/etm.2019.7533

Table III.

Parameters of liquid chromatography-tandem mass spectrometry method for measuring glutamate, aspartate, glutamine and asparagine.

Parameter Value
Mobile phase Water containing 0.05% (v/v) formic acid
Column Phenomenex Kinetex F5 column (4.6×100 mm, 2.6 µm)
Column temperature (°C) 30
Flow rate (ml/min) 0.3
Capillary voltage (kV) 3.5
Drying gas temperature (°C) 350
Drying gas flow (l/min) 10
Nebulizer pressure (psi) 40
MRM transition (m/z)
  Glu 148→84
  Gln 147→83
  Asp 134→88
  Asn 133→87
  Glu-IS 151→87
  Asp-IS 137→91
Dwell (msec)
  Glu 100
  Gln 100
  Asp 100
  Asn 100
  Glu-IS 100
  Asp-IS 100
Fragmentor (V)
  Glu 80
  Gln 80
  Asp 50
  Asn 50
  Glu-IS 80
  Asp-IS 50
CE (eV)
  Glu 18
  Gln 5
  Asp 7
  Asn 4
  Glu-IS 18
  Asp-IS 7

Glu-IS, 2H3-Glutamate; Asp-IS, 2H3-Aspartate; CE, Collision Energy