Table 3.

Comparison of total beam-on time in minutes per patient SRS plan, assuming a dose rate of 1,400 MU/min for the linac-based options.

Beam-On Time (min)
Gamma Knife Elements HyperArc-VMAT Manual-VMATA Manual-VMATB
158 4.09 5.95 3.80 9.99
149 5.27 7.08 3.14 7.28
97.7 5.63 4.49 4.71 9.13
182 5.77 6.53 7.32 9.61
128 5.96 6.41 4.85 7.47
156 5.39 5.87 4.13 7.24
143 4.61 5.35 4.95 8.41
162 5.24 6.86 4.74 8.29
160 4.75 6.73 4.29 7.96
166 6.08 6.53 6.08 7.60
111 3.17 4.34 3.55 6.52
116 3.88 6.84 3.43 7.48
146 3.96 5.16 3.05 6.36
111 4.22 6.07 4.04 7.83
113 4.27 7.39 4.85 10.1
141 4.52 5.60 4.82 7.06
140.0 ± 24.4 4.80 ± 0.849 6.08 ± 0.897 4.48 ± 1.09 8.02 ± 1.16

The last row lists the mean ± standard deviation across all 16 patients, per SRS treatment technique.