Table 4.

Details on the ICD programming divergences from the recommended settings (presented as Figures 5 and 6 in Appendix)

TOTAL
(n = 185)
SC-ICD
(n = 44)
DC-ICD
(n = 48)
CRT-D
(n = 93)
Primary prevention (pts) 121 19 25 77
 FVT detection- ON [n, (%)] 19 (10.2) 1 (2.3) 4 (8.3) 14 (15.1%)
 VT detection- ON [n, (%)] 90 (74.4) 15 (78.9) 18 (72.0) 57 (64.0)
 PR Logic/ Wavelet- OFF or Monitor [n, (%)] 18 (14.9) 3 (15.8) 2 (8.0) 13 (16.9)
 Onset ON or OFF [n, (%)] 49 (40.5) 12 (53.1) 9 (36.0) 28 (36.4)
 T-Wave- OFF [n, (%)] 6 (5) 3 (15.8) 0 3 (3.9)
 RV-Lead Noise [n, (%)] 3 (2.5) 2 (10.5) 0 1 (1.3)
 ATP before charging 8 (6.6) 1 (5.3) 2 (8.0) 5 (6.5)
 FVT- ATP & shocks 121 (100) 19 (100) 25 (100) 77 (100)
 VT- ATP & shocks 121 (100) 19 (100) 25 (100) 77 (100)
Secondary prevention (pts) 64 25 23 16
 FVT detection- ON [n, (%)] 19 (29.7) 8 (32.0) 9 (39.1) 2 (12.6)
 VT detection- OFF [n, (%)] 11 (17.2) 6 (24.0) 5 (21.7) 0
 PR Logic/ Wavelet- OFF or Monitor [n, (%)] 4 (6.3) 3 (12.0) 0 1 (6.3)
 Onset ON or OFF [n, (%)] 32 (50.0) 10 (40.0) 14 (60.8) 8 (50.0)
 ATP before charging 8 (12.5) 3 (12.0) 2 (8.7) 3 (18.8)
 FVT- ATP & shocks 121 (100) 19 (100) 25 (100) 77 (100)