. 2019 Apr-Jun;44(2):126–134. doi: 10.4103/jmp.JMP_118_18

Table 2.

Conformity index for the task group 119 test cases from the analytical anisotropic algorithm and Acuros XB algorithm-calculated RapidArc plans for the Clinac-iX (6 MV) and TB-STx (6MV_filtered beam and 6MV_flattening filter-free beam), respectively

Structure CI

Clinac-iX (6MV FB) TB-STx (6MV FB) TB-STx (6MV FFFB)



AAA AXB AAA AXB AAA AXB
C Shape (easier) 1.037 0.970 1.035 0.953 1.033 0.785
C Shape (hard) 1.055 0.997 1.046 0.974 1.049 0.911
H and N 0.923 0.955 0.929 0.492 0.935 0.399
Multi-target 1.045 0.963 1.069 1.048 1.077 1.010
Prostate 0.989 0.943 0.977 0.923 0.971 0.851
Mean±SD 1.010±0.055 0.966±0.020 1.011±0.057 0.878±0.221 1.013±0.058 0.791±0.234

TB: TrueBeam, FB: Filtered beam, FFFB: Flattening filter-free beam, CI: Conformity index, AAA: Analytical anisotropic algorithm, AXB: Acuros XB, SD: Standard deviation