Skip to main content
. 2019 Apr-Jun;44(2):126–134. doi: 10.4103/jmp.JMP_118_18

Table 4.

Dose-spillage volumes normalized to target volume, along with their P values comparing for different multileaf collimator and beam type using analytical anisotropic algorithm, Acuros XB calculation for Clinac-iX (6 MV) and TB-STx (6MV_filtered beam and 6MV_falttening filter-free beam), respectively

Dose spillage volume Millennium-120 (FB) P HD-120 (FB) P HD-120 (FFFB) P AAA P
Mill
versus HD
AXB P
Mill
versus HD
AAA P
FB
versus FFFB
AXB P
FB
versus FFFB



AAA AXB AAA AXB AAA AXB
VHS 0.11±0.05 0.09±0.04 0.004 0.11±0.06 0.08±0.04 0.028 0.10±0.06 0.06±0.04 0.007 0.828 0.034 0.485 0.327
VIS 2.54±1.03 2.44±0.99 0.005 2.46±1.09 2.27±0.99 0.010 2.44±1.08 2.22±0.98 0.011 0.171 0.021 0.653 0.357
VLS 12.91±4.51 11.84±4.34 0.019 11.53±4.63 10.95±4.36 0.016 11.61±4.59 10.97±4.36 0.019 0.026 0.010 0.535 0.904