. 2019 Apr-Jun;44(2):126–134. doi: 10.4103/jmp.JMP_118_18

Table 5.

Gamma analysis of the task group 119 test cases from the analytical anisotropic algorithm and Acuros XB algorithm-calculated RapidArc plans for the Clinac-iX (6 MV) and TB-STx (6MV_filtered beam and 6MV_flattening filter-free beam), respectively

Structure Gamma (3%/3mm TH: 10%)

Clinac-iX (6MV FB) TB-STx (6MV FB) TB-STx (6MV FFFB)



AAA AXB AAA AXB AAA AXB
C shape (easier) 98.1 93.9 98.1 95.1 98.6 93.2
C shape (hard) 98.5 96.6 98.7 96.4 99.2 95.3
H and N 99.1 93.5 99.9 98.3 99.5 94.4
Multi-target 97.7 94.1 96.6 95.4 94.6 93.5
Prostate 99.0 97.4 99.4 95.2 98.8 93.4
Mean±SD 98.5±0.6 95.1±1.8 98.5±1.3 96.1±1.3 98.1±2.0 94.0±0.9

TH: Threshold, FB: Filtered beam, FFF: Flattening filter free, AAA: Analytical anisotropic algorithm, AXB: Acuros XB, SD: Standard deviation