. 2019 Apr-Jun;44(2):126–134. doi: 10.4103/jmp.JMP_118_18

Table 6.

Ion chamber measurements in comparison to the treatment planning system-calculated dose at the ArcCHECK isocenter for the task group 119 test cases from the analytical anisotropic algorithm and Acuros XB algorithm-calculated RapidArc plans for the Clinac-iX (6 MV) and TB-STx (6MV_filtered beam and 6MV_flattening filter-free beam), respectively

Structure IC versus TPS calculation (% difference)

Clinac-iX (6MV FB) TB-STx (6MV FB) TB-STx (6MV FFFB)



AAA AXB AAA AXB AAA AXB
C shape (easier) 3.4 4.7 4.1 4.6 4.4 4.9
C shape (hard) −2.7 2.5 2.2 2.3 3.2 3.6
H and N 2.4 2.6 1.6 3.5 −0.7 1.7
Multitarget 1.4 1.2 1.3 1.1 −2.0 −0.3
Prostate 2.8 0.5 4.1 4.6 2.0 1.9
Mean±SD 1.5±2.5 2.3±1.6 2.7±1.4 3.2±1.5 1.4±2.7 2.3±2.0

TPS: Treatment planning system, IC: Ion Chamber, TB: TrueBeam, FFF: Flattening filter free, AAA: Analytical anisotropic algorithm, AXB: Acuros XB, SD: Standard deviation