Table 4. Parameters for capillary electrophoresis for each group.

Group CE Parameters
1, 9, 12 ABI 3500xl, ABI3130xl, 36cm array using POP-4: oven temp (60 °C), run time (1500 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (15 sec), injection voltage (1.2 kV), and data delay (1 sec).
11 ABI 3500xl 50cm array with POP-7: oven temp (60 °C), run time (2000 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (15 sec), injection voltage (1.2 kV), and data delay (1 sec).
2 ABI 3130xl 50cm array with POP-7: oven temp (60 °C), run time (6000 sec), run voltage (15kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (15 sec), injection voltage (1.6 kV), and data delay (200 sec).
4 ABI 3130xl 50cm array with POP-4: oven temp (60 °C), run time (2200 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (15 sec), injection voltage (1.6 kV), and data delay (500 sec).
3, 6 ABI 3130xl 36cm array with POP-4: oven temp (60 °C), run time (1500–2700 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (10 sec), injection voltage (3 kV), and data delay (1 sec).
7 ABI 3730xl 50cm array with POP-7: oven temp (63 °C), run time (5000 sec), run voltage (10 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (25 sec), injection voltage (1.6 kV), and data delay (NA).
8 ABI 3730xl 50cm array with POP-7: oven temp (63 °C), run time (2100 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (10 sec), injection voltage (1.6 kV), and data delay (500 sec).
5 ABI 3730xl 50cm array with POP-7: oven temp (60 °C), run time (3000 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (15 sec), injection voltage (1.2 kV), and data delay (1 sec).
10 ABI 3730 36cm array with POP-7: oven temp (66 °C), run time (1200 sec), run voltage (15 kV), pre-run time (180 sec), pre-run voltage (15 kV), injection time (10 sec), injection voltage (2 kV), and data delay (120 sec).