Skip to main content
. 2019 Feb 19;33(4):e22856. doi: 10.1002/jcla.22856

Table 2.

Equations produced by curve fits using GraphPad Prism 7 and the HI thresholds calculated from the equations

Analytes TAE Functions R2 Hemolysis Index threshold Hemoglobin † (g/L)
Aspartate aminotransferase +20% Bias% = 13.63 + 0.63*HI‐0.29e‐03*HI^2 0.98 P < 0.05 10.16 0.65
Total bile acid −10%a Bias% = 0.65‐0.26*HI+18.53e‐05*HI^2 0.98 P < 0.05 42.23 0.92
Lactate dehydrogenase +20% Bias% = 24.04 + 1.47*HI‐91.81e‐05*HI^2 0.98 P < 0.05 NA >0.5
Creatine kinase +30% Bias% = −2.03 + 0.41*HI‐0.10e‐02*HI^2 + 1.04e‐06* HI^3 0.99 P < 0.05 99.96 1.40
Hydroxybutyrate Dehydrogenase +30% Bias% = 18.15 + 1.50*HI‐92.35e‐05*HI^2 0.99 P < 0.05 7.94 0.63
Uric acid −17% Bias% = −0.06*HI+5.20e‐05*HI^2 0.85 P < 0.05 500 4.73
Creatine kinase‐MB +3 s
(11.58 U/L)
Bias = −1.11 + 0.29*HI‐21.21e‐05*HI^2 + 1.75e‐007*HI^3 0.99 P < 0.05 45.20 0.94
Potassium +0.5
mmol/L
Bias = 0.82*log(HI) 0.90 P < 0.05 4.07 0.60
Sodium −4
mmol/L
Bias = −94.42e‐04*HI+0.74 0.89 P < 0.05 501.8 4.75

TAE, total allowed error recommended by CLIA’88 regulations; NA, not available. The total allowed error for hydroxybutyrate dehydrogenase was assumed to be 30% according to experience.

a

±10% bias was set as the accepted TAE for the analytes because they are not included in the CLIA’88 regulations. Hemoglobin was estimated from equation HI = 120*Hb‐68, supplement Figure S1.