Skip to main content
. 2019 May 13;92(1098):20180670. doi: 10.1259/bjr.20180670

Table 5.

Different demands of MRI acquired for diagnostic and radiotherapy purposes in cervix and rectal cancer

MRI for diagnosis MRI for radiotherapy
Couch Soft, often concave
Maximized for patient comfort
Needs to be flat, the same as in RT delivery
Patient positioning Comfortable
Supine
As for RT delivery
Supine
Immobilization devices None Combifix knee support to stabilize pelvis
Bowel artefact management IM Buscopan
Anterior abdominal wall compression
Saturation bands
IM Buscopan may be used in MRI simulation but may not be acceptable during daily treatment within MRI treatment workflow
Bladder status Empty Full
Coil placement Pelvic coil centred on tumour Anterior coil supports prevent distortion of external body contour
Customized MR simulators may incorporate posterior coils into a flat couch
Field strength Increasing strength improves signal to noise, but is more expensive and requires more room Increasing field strength increases geometric distortion
Coverage High resolution FOV limited to tumour High resolution FOV must encompass entire tumour target
Sequences including external body contour required for dose calculation
Preferred
sequence
2d T 2W high resolution at tumour with ≤3 mm slice thickness, and voxel size <1 mm
Imaging plane perpendicular to the rectum or cervical canal
T 2W 3d < 1 mm isotropic voxel size for target delineation
Imaging plane true axial acquired perpendicular to the system
Geometric accuracy Less important Essential to localize the target
Electron density/ material composition information Not required Not required in a CT/ MRI combined workflow, but essential in MR-only simulation and MR treatment workflow

FOV, field of view;RT, radiotherapy.