Skip to main content
. 2019 Apr 9;92(1098):20180075. doi: 10.1259/bjr.20180075

Table 1.

Multiparametric MRI protocols for Home and 2 Referral Centres

Parameter Centre Axial T2 TSE Sagittal T2 TSE Coronal T2 TSE DWI DCE
TR (ms) Home 3960 4560 3780 4975 3.3
1 4500 6540 4300 5275 3.9
2 3420 4610 2870 5600 -
TE (ms) Home 120 120 120 82 1.6
1 106.7 88.7 102.2 84.4 1.6
2 99 103 103 89 -
Field of view (mm) Home 200 × 200 240 × 240 200 × 200 340 × 340 230 × 230
1 230 × 230 230 × 230 230 × 230 340 × 340 300 × 300
2 250 × 250 220 × 220 200 × 200 267 × 267 -
Matrix size Home 288 × 288 384 × 384 288 × 288 98 × 128 160 × 160
1 288 × 256 288 × 224 320 × 256 98 × 128 192 × 192
2 102 × 160 160 × 320 192 × 320 102 × 160 -
Slice thickness (mm) Home 3 3 3 4 4 (2mm overlap)
1 3 3 3 4 5
2 3 3 3 3.6 -
Flip angle (⁰) Home 90 90 90 180 12
1 90 90 90 90 12
2 150 150 150 - -
Scan time (mins) Home 5:09 3:16 4:55 5:03 5:38
1 3:41 4:41 3:31 5:16 6:05
2 4:40 2:00 3:00 9:38 -
Temporal resolution (s) Home - - - - 12
1 - - - - 12
2 - - - - -