Diagnostic accuracy of UC‐3500 vs reference methods for clinically most relevant parameters of urinalysis

Parameter Test strip No. of Samples Sensitivity (%) Specificity (%) NPVa (%) PPVb (%)
Glucose UC‐11A 59 100 60 100 88
Protein UC‐9A 67 94.2 82.4 82.4 94.2
UC‐11A 67 94.2 88.2 83.3 96.1
Hemoglobin peroxidase UC‐9A 347 85.1 89.3 96.2 65.5
UC‐11A 347 85.1 88.6 96.1 64.0
Leukocyte esterase UC‐9A 347 80.5 91.7 93.8 75.0
UC‐11A 347 81.7 92.8 94.6 77.9
Albumin UC‐11A 177 81.8 89.2 89.1 81.8