Skip to main content
. 2019 Jun 25;57(7):e01604-18. doi: 10.1128/JCM.01604-18

TABLE 3.

Detection of C. auris in spiked clinical specimens on the BD Maxa

CFU/250 μl specimen Mean CT ± SD for:
Saline Serum Urine Sputum Wound
5E + 02 26.93 ± 0.68 28.83 ± 1.63 25.15 ± 0.93 26.35 ± 0.92 26.78 ± 1.56
5E + 01 30.22 ± 1.65 32.55 ± 1.06 28.56 ± 1.72 29.05 ± 1.77 28.20 ± 0.14
5E + 00 33.27 ± 1.33 NDb 31.10 ± 3.49 32.70 ± 1.41 30.90 ± 2.05
a

C. auris was amplified by real-time PCR in spiked clinical specimen in 3 independent experiments.

b

ND, not detected; there was no detection in unspiked specimen control.