. 2019 Mar 7;35(7):1327–1337. doi: 10.1007/s10554-019-01574-0

Table 2.

CRT implantation and follow up characteristics

Target group (n = 6) Non-target group (n = 9) p-value
Distance to target sites
 Distance to target (mm) 8 [0–22] 26 [17–46] 0.04
 Distance to infarct (mm) 22 [21–23] (n = 2) 26 [14–51] 0.51
 Distance to left phrenic nerve (mm) 44 [18–54] 44 [36–n/a] 0.61
Implantation characteristics
 Implantation duration (min) 146 ± 38 127 ± 35 0.38
 LV lead implantation duration (min) 47 ± 18 55 ± 28 0.57
 Fluoroscopy time (min) 36 ± 15 28 ± 12 0.30
 Total radiation dose (cGycm2) 6758 ± 4201 8242 ± 6446 0.70
 Pre-procedural 3D-angiogram radiation (cGycm2) 1188 ± 262 1449 ± 452 0.41
 Post-procedural 3D-angiogram radiation (cGycm2) 1313 ± 333 1491 ± 439 0.57
 Radiation dose CRT only (cGycm2) 5753 ± 3038 5303 ± 5847 0.91
LV lead electrical properties
 Paced QRS duration (ms) 153 ± 22 170 ± 22 0.18
 Decrease QRS duration (ms) − 12 ± 13 − 9 ± 27 0.10
 Pacing threshold (V) 0.65 ± 0.39 0.58 ± 0.20 0.64
 QLV (ms) 150 ± 8 130 ± 30 0.23
 Ratio QLV/QRS (%) 85 ± 10 81 ± 16 0.66
Echocardiographic follow up
 LV end-systolic volume change (%) − 30 ± 10 − 19 ± 19 0.28
 LV ejection fraction change from baseline (%) 15 ± 5 10 ± 12 0.30

Values are in mean ± SD, median [interquartile range]

LV left ventricular, QLV interval from Q on the surface ECG to local sensing at the LV electrogram