Table 3.
Summary of the certification results for the SRM 3389 and MIGS-1 solutions.
| Gravimetry |
LC/UVa |
LC/MSc |
|||
|---|---|---|---|---|---|
| Ginsenoside | Mass Fraction (mg/g) |
Mass Fractionb (mg/g) |
RSD (%) |
Mass Fractionb (mg/g) |
RSD (%) |
| SRM 3389 | |||||
| Rg1 | 0.51 | 0.518 ± 0.013 | 2.53 | 0.530 ± 0.023 | 4.27 |
| Re | 1.53 | 1.55 ± 0.039 | 2.52 | 1.54 ± 0.095 | 6.13 |
| Rf | 1.27 | 1.31 ± 0.039 | 3.02 | 1.31 ± 0.138 | 10.59 |
| Rb1 | 3.81 | 3.80 ± 0.097 | 2.55 | 3.86 ± 0.345 | 8.92 |
| Rc | 0.76 | 0.761 ± 0.019 | 2.55 | 0.750 ± 0.059 | 7.81 |
| Rb2 | 0.89 | 0.903 ± 0.023 | 2.54 | 0.877 ± 0.078 | 8.85 |
| Rd | 0.25 | 0.251 ± 0.008 | 3.18 | 0.246 ± 0.025 | 8.68 |
| MIGS-1: Control | |||||
| Rg1 | 0.258 ± 0.005 | 0.254 ± 0.005 | 1.90 | 0.257 ± 0.015 | 5.67 |
| Re | 0.506 ± 0.002 | 0.504 ± 0.009 | 1.86 | 0.525 ± 0.028 | 5.40 |
| Rf | - | - | - | - | - |
| Rb1 | 1.249 ± 0.016 | 1.24 ± 0.024 | 1.96 | 1.35 ± 0.114 | 8.42 |
| Rc | 0.505 ± 0.005 | 0.489 ± 0.010 | 1.98 | 0.497 ± 0.044 | 8.89 |
| Rb2 | 0.252 ± 0.002 | 0.254 ± 0.005 | 2.02 | 0.261 ± 0.017 | 6.39 |
| Rd | 0.507 ± 0.006 | 0.489 ± 0.010 | 1.91 | 0.504 ± 0.055 | 10.92 |
LC/UV values quantified using an external calibration approach.
The mass fraction values are the average ± standard deviation of ten measurements for each calibrant, candidate SRM 3389, and MIGS-1 control solutions. In the cases of LC/MS, the candidate SRM 3389 was measured nine times.
LC/MS values quantified using ginsenoside Rh1 for the internal standard calibration approach.