Published in final edited form as: J Chromatogr A. 2018 Sep 7;1574:114–121. doi: 10.1016/j.chroma.2018.09.011

Table 3.

Summary of the certification results for the SRM 3389 and MIGS-1 solutions.

Gravimetry
LC/UVa
LC/MSc
Ginsenoside Mass Fraction
(mg/g)
Mass
Fractionb
(mg/g)
RSD
(%)
Mass
Fractionb
(mg/g)
RSD
(%)
SRM 3389
Rg1 0.51 0.518 ± 0.013 2.53 0.530 ± 0.023 4.27
Re 1.53 1.55 ± 0.039 2.52 1.54 ± 0.095 6.13
Rf 1.27 1.31 ± 0.039 3.02 1.31 ± 0.138 10.59
Rb1 3.81 3.80 ± 0.097 2.55 3.86 ± 0.345 8.92
Rc 0.76 0.761 ± 0.019 2.55 0.750 ± 0.059 7.81
Rb2 0.89 0.903 ± 0.023 2.54 0.877 ± 0.078 8.85
Rd 0.25 0.251 ± 0.008 3.18 0.246 ± 0.025 8.68
MIGS-1: Control
Rg1 0.258 ± 0.005 0.254 ± 0.005 1.90 0.257 ± 0.015 5.67
Re 0.506 ± 0.002 0.504 ± 0.009 1.86 0.525 ± 0.028 5.40
Rf - - - - -
Rb1 1.249 ± 0.016 1.24 ± 0.024 1.96 1.35 ± 0.114 8.42
Rc 0.505 ± 0.005 0.489 ± 0.010 1.98 0.497 ± 0.044 8.89
Rb2 0.252 ± 0.002 0.254 ± 0.005 2.02 0.261 ± 0.017 6.39
Rd 0.507 ± 0.006 0.489 ± 0.010 1.91 0.504 ± 0.055 10.92
a

LC/UV values quantified using an external calibration approach.

b

The mass fraction values are the average ± standard deviation of ten measurements for each calibrant, candidate SRM 3389, and MIGS-1 control solutions. In the cases of LC/MS, the candidate SRM 3389 was measured nine times.

c

LC/MS values quantified using ginsenoside Rh1 for the internal standard calibration approach.