Table 1.

Summarizes the main data drawn from previous reports and from our case

S-ICD indication LVAD type Operating rate (rpm) Cycles/s (Hz) Pump placement EMI Usefulness S-ICD
Saeed et al.6 Before LVAD NIDCM HeartWare HVAD® 2500 41.66 Thorax Yes No
Ahmed et al.11 Before LVAD NIDCM HeartWare HVAD® 2800 46.67 Thorax Yes No
Gupta et al.7 After LVAD T-ICD extraction HeartMate II™ 9800 163.33 Abdomen No Yes
Raman et al.8 After LVAD NIDCM NSVT HeartMate II™ 6000–10 000a 100–166.66 Abdomen No Yes
Migliore et al.10 Before LVAD T-ICD extraction Jarvik 2000® NA (7000–13 000)a 116.66–216.66 Thorax No Yes
Pfeffer et al.9 Before LVAD NIDCM HeartMate 3™ NA (2000–5500)a 33.33–91.66 Thorax Yes No
Saini et al.12 Before LVAD NIDCM HeartMate 3™ NA (2000–5500)a 33.33–91.66 Thorax Yes No
López-Gil et al. (this manuscript) Before LVAD NIDC NSVT HeartMate 3™ 5100 85 Thorax Yes No
a

Parenthesis values are operating rate range of the specific LVAD when not specified in the report.

After LVAD, the S-ICD was implanted after the LVAD; Before LVAD, the S-ICD was implanted before the LVAD; EMI, electromagnetic interference; Hz, rpm/60; LVAD, left ventricular assist device; NA, non-available; NIDCM, non-ischaemic dilated cardiomyopathy; NSVT, non-sustained ventricular tachycardia; S-ICD, subcutaneous implantable cardiac defibrillator; T-ICD, transvenous ICD.