Table 1.
Summarizes the main data drawn from previous reports and from our case
S-ICD indication | LVAD type | Operating rate (rpm) | Cycles/s (Hz) | Pump placement | EMI | Usefulness S-ICD | |
---|---|---|---|---|---|---|---|
Saeed et al.6 | Before LVAD NIDCM | HeartWare HVAD® | 2500 | 41.66 | Thorax | Yes | No |
Ahmed et al.11 | Before LVAD NIDCM | HeartWare HVAD® | 2800 | 46.67 | Thorax | Yes | No |
Gupta et al.7 | After LVAD T-ICD extraction | HeartMate II™ | 9800 | 163.33 | Abdomen | No | Yes |
Raman et al.8 | After LVAD NIDCM NSVT | HeartMate II™ | 6000–10 000a | 100–166.66 | Abdomen | No | Yes |
Migliore et al.10 | Before LVAD T-ICD extraction | Jarvik 2000® | NA (7000–13 000)a | 116.66–216.66 | Thorax | No | Yes |
Pfeffer et al.9 | Before LVAD NIDCM | HeartMate 3™ | NA (2000–5500)a | 33.33–91.66 | Thorax | Yes | No |
Saini et al.12 | Before LVAD NIDCM | HeartMate 3™ | NA (2000–5500)a | 33.33–91.66 | Thorax | Yes | No |
López-Gil et al. (this manuscript) | Before LVAD NIDC NSVT | HeartMate 3™ | 5100 | 85 | Thorax | Yes | No |
Parenthesis values are operating rate range of the specific LVAD when not specified in the report.
After LVAD, the S-ICD was implanted after the LVAD; Before LVAD, the S-ICD was implanted before the LVAD; EMI, electromagnetic interference; Hz, rpm/60; LVAD, left ventricular assist device; NA, non-available; NIDCM, non-ischaemic dilated cardiomyopathy; NSVT, non-sustained ventricular tachycardia; S-ICD, subcutaneous implantable cardiac defibrillator; T-ICD, transvenous ICD.