Skip to main content
. 2019 Jun 28;8(3):61. doi: 10.1167/tvst.8.3.61

Table 3.

Statistical Distributions of CNR Values and Paired t-Test Result for Both Systems

Parameter
Low-Cost OCT
Heidelberg Spectralis
Mean CNR ± std error 1.592 ± 0.021 1.687 ± 0.027
95% Confidence interval 1.551–1.634 1.634–1.740
Range 0.984–2.232 0.877–2.546
Mean CNR difference ± std error 0.09 ± 0.02
Paired t-test P value <0.0001