Fig. 2.

Fig. 2

System resolution calibration using a chromium grating coated on a coverslip. (a) Fitted edge spread function (ESF) and line spread function (LSF) of experimental data shows that PAM has a lateral resolution of 4.1 μm. (b) Summation of two A-line axial spread function shows that PAM has an axial resolution of 37.0 μm. (c) Fitted ESF and LSF of experimental data show that OCT has a lateral resolution of 3.8 μm at −6 dB. (d) A-line spread function shows that OCT has an axial resolution of 4.0 μm at −6 dB in air.