1 |
Fan folded geometry (1 × 1 × 1 cm3) |
PSI‐5A 4E piezo sheets |
2.18 µW |
1 µW |
Too high natural frequency and impractical |
20
|
2 |
Fan folded geometry (2 × 0.5 × 1 cm3) |
PZT |
>10 µW |
Leadless pacemakers need <10 µW |
Proof mass of 21 g makes harvester too heavy and infeasible. |
21
|
3 |
Cantilever beam of 0.032 m span length |
PVDF |
≃25 V |
5–10 µW |
Geometrical constraints are not satisfied. |
22
|
4 |
Sheet of 8.4 µm thickness |
PMN‐PT |
145 µA, 8.2 V |
100 µA, 3 V |
Stimulator was readily implemented on live rat. |
23
|
5 |
Zig‐zag beam of length 27 mm length |
PZT‐5A |
10 µW |
N/A |
Too long for intravenous cavity |
12
|
6 |
Thin rectangular film on plastic substrate |
PIMNT |
V
oc = 11 V |
N/A |
Self‐powered deep brain simulation for a live animal |
24
|
|
|
|
I
cc = 285 µA |
|
|
|