Skip to main content
. 2019 May 10;6(Pt 4):586–602. doi: 10.1107/S2052252519005128

Table 1. Estimated attenuation contributions by the background absorbers using FFAST theoretical data of X-ray mass attenuation coefficients, and the geometry of the experimental setup.

For fluorescence geometry one needs to separate the incident flux at incident energy through the upstream path lengths, from the fluorescent emission path lengths, hence the parameters displayed in the table. The path lengths were determined from the measured geometry of the experimental set-up.

Background absorber Chemical formula Density (g cm−3) Path length (cm) t)nominal (g cm−2) (μ/ρ)FFAST (cm2 g−1) (μ/ρ)(ρt)FFAST
Upstream, incident E, e.g. 8.45 keV
Air N2 (78%) + O2 (21%) + Ar (0.93%) 0.0011 28.9 (6) 0.03179 7.696 0.2447
Detector gas N2 0.0012 19.0 (1) 0.02280 6.171 0.1407
Kapton (polymide) C12H10N2O5 1.42 0.010 (1) 0.0142 5.610 0.0797
Silicone (adhesive) CH3—Si2O2—C4H9 0.968 0.0060 (6) 0.00581 22.29 0.1295
He gas (cryostat) He 0.0001785 1.1 (1) 0.000196 0.2633 0.0000516
Downstream, fluorescent E, Ni Kα 7.39 keV
Air N2 (78%) + O2 (21%) + Ar (0.93%) 0.0011 8.2 (1) 0.00902 11.47 0.1035
Kapton (polymide) C12H10N2O5 1.42 0.0060 (6) 0.00852 8.305 0.0708
Silicone (adhesive) CH3—Si2O2—C4H9 0.968 0.0060 (6) 0.00581 32.59 0.1893
He gas (cryostat) He 0.0001785 1.1 (1) 0.000196 0.2932 0.0000575