Optical coherence tomography analysis of coronary lesions at the 9-month follow-up (malapposition)

Time Parameters OCT
DEB-BMS (n = 8) DES (n = 8) p value
Post-procedure Mean number of struts/mm section 13.8 ± 1.6 12.7 ± 0.9 0.098
Mean number of malapposed struts/mm section 1.2 ± 0.7 0.9 ± 1.0 0.458
% malapposed struts/mm section 9.5 ± 5.1 7.0 ± 7.6 0.541
At 9 months Mean number of struts/mm section 13.3 ± 1.7 12.3 ± 2.6 0.374
Mean number of uncovered struts/mm section 10.8 ± 28.0 8.1 ± 8.9 0.794
Mean number of malapposed struts/mm section 4.6 ± 12.3 2.5 ± 6.0 0.666
% of struts uncovered/mm section 4.7 ± 12.2 3.0 ± 3.3 0.716
% of struts malapposed/mm section 1.99 ± 5.37 0.9 ± 2.2 0.597

Values are presented as mean ± SD.

OCT, optical coherence tomography; DEB, drug-eluting balloon; BMS, bare metal stent; DES, drug-eluting stent.