Skip to main content
. 2019 Jun 17;20(7):109–120. doi: 10.1002/acm2.12663

Table 4.

Variations in ITVs for AIP and MIP CT datasets in the irregular motion patterns for gated and nongated plans.

Phase range Respiratory motion pattern Image type ITV (cm3) ITV variation (AIP/MIP)
Gated 30%–70% Slightly irregular AIP 8.36 −7.73%
MIP 9.06
Amplitude irregular AIP 7.28 −7.61%
MIP 7.88
Period irregular AIP 10.12 −6.73%
MIP 10.85
Period and amplitude irregular AIP 7.33 −6.15%
MIP 7.81
Nongated 10%–90% Slightly irregular AIP 10.12 −12.31%
MIP 11.54
Amplitude irregular AIP 9.18 −12.99%
MIP 10.55
Period irregular AIP 10.2 −14.00%
MIP 11.86
Period and amplitude irregular AIP 8.19 −15.22%
MIP 9.66

AIP, average intensity projection; ITV, internal target volumes; MIP, maximum intensity projection.