Skip to main content
. 2019 Jun 1;9(3):345–354. doi: 10.31661/jbpe.v0i0.1047

Table 2.

Percentage of dose increment and reduction for different men and women prosthesis.

Material Percentage of dose increment (%) Percentage of dose reduction (%)
Men Women Men Women
Cr _Co _ Mo alloy 6.81 6.4 41.18 40.87
SST 6.1 5.96 40.66 40.36
Ti alloy 4 3.87 37.76 36.94
Ti 4.29 3.94 37.51 36.69