TABLE 1.

Magnetic resonance imaging parameters for the gel phantom measurements.

Sequence (position in QA protocol) Localizer (1) T1 (2) Bold sensitive EPI (3, 6) Diffusion sensitive EPI (4) Bold sensitive EPI (5)
Repetition time (TR) 8.6 ms 1900 ms 2000 ms 3900 ms 177 ms
Echo time (TE) 4 ms 2.26 ms 30 ms 90 ms 30 ms
Field of view (FoV) 250 mm 256 mm 210 mm 256 mm 210 mm
Matrix size 256 × 256 256 × 256 64 × 64 128 × 128 64 × 64
Slice thickness 7.0 mm 1.0 mm 3.0 mm 2.0 mm 3.0 mm
Distance factor 50% 20% 0% 20%
Flip angle 20° 70° 70°
Phase encoding direction Anterior >> posterior, anterior >> posterior, right >> left Anterior >> posterior Anterior >> posterior Anterior >> posterior Anterior >> posterior
Bandwidth (Hz/Px) 320 200 2894 1502 2894
Acquisition order (series) Sequential (interleaved) Single shot (ascending) Interleaved (interleaved) Interleaved (interleaved) Interleaved (interleaved)
Number of slices Two in each direction 176 34 30 3
Measurements 1 1 200 322
Effective voxel size (mm) 1.0 × 1.0 × 7.0 1.0 × 1.0 × 1.0 3.3 × 3.3 × 3.0 2.0 × 2.0 × 2.0 3.3 × 3.3 × 3.0
Acquisition time (TA) 0.25 4:26 6:44 2:14 1:00