Abstract
We present a case with imaging artefacts on mammography and Magnetic Resonance Imaging (MRI) caused by iron oxide particles. After being diagnosed with the medullary cancer of the breast, the female patient had a breast conserving surgery on right breast. Iron oxide particles were used for the detection of the sentinel lymph node during operation. On follow ups, a de novo density on mammography, which was initially thought to be a new tumour, was found. MR images proved that the lesion is an artefact caused by iron oxide accumulation. Our aim in this case study is to underline and discuss the imaging artefacts caused by these particles and raise awareness.
Keywords: Breast, iron oxide, mammography, MRI
Introduction
A sentinel lymph node biopsy (SLNB) is a procedure during which the sentinel lymph node, which is the main drainage pathway, is identified, excised, and evaluated histopathologically to detect cancer cells. A negative SLNB result suggests that cancer cells have not developed the ability to spread to nearby lymph nodes or other organs. Sentinel lymph node biopsy (SLNB) is a feasible and reliable method for staging the axilla before breast cancer surgery (1–6). The conventional method is with radiotracer 99mtc using blue dye. In the latest years a new method for detecting sentinel node called iron oxide particle technique (“SentiMag® technique”), has been developed as an alternative to the radiotracer. In this new method a liquid which contains iron oxide particles is injected subareolarly. The liquid dissolves in the breast towards the axilla by manual massage. Later, a probe is used to detect iron oxide particles with ferromagnetic effect. Prospective clinical studies like “The Central-European SentiMag study” and meta analyses of earlier studies showed that magnetic SLNB method can be performed easily, safely and is a promising alternative to the radioactive method. (7–10). By time these iron oxide particles disseminate and are cleared through the lymphatics. However, we know that in some cases they might stay in the breast tissue for years and reduce imaging quality and sensitivity by causing artefacts (7, 8). Our aim in this case study is to discuss those imaging artefacts and help radiologists be aware of them.
Case Presentation
We present a 41-year-old female patient who had breast conserving surgery for biopsy proven medullary cancer a year ago. Informed consent was taken from the patient. She had a 7×3 mm sized palpable right breast mass. Although being palpable, it was occult on mammography (Figure 1a). The lesion was detected by sonography. On sonography a hypoechoic lesion with well-defined borders was found and classified as category 3 by Breast Imaging Reporting and Data System (BI-RADS) (Figure 1b). However, the patient insisted on biopsy and histopathological diagnosis revealed the medullary cancer. A breast conserving surgery was performed on right breast. 12 months later the patient was referred to mammography unit for routine follow up. On mammography, a spiculated de novo lesion that was denser than the radiographically dense breast parenchyma was identified on right upper breast (Figure 2a). The lesion was first thought to be compatible with recurrence. A new primary was also considered since the previous lesion was occult on mammography. We could not identify the new lesion on sonography. Magnetic Resonance images were obtained to delineate the whole breast. On MRI, a marked magnetic susceptibility artefact obscuring the central part of the breast was seen. These areas were attributed to previous iron oxide particles which were injected for SLN detection (Figure 2b, c) (SentiMag® technique).
Figure 1. a, b.
Right MLO Mammography images. No mass is seen (a). A hypoechoic lesion with well-defined borders is seen on sonography compatible with BI-RADS 3 lesion (b)
Figure 2. a–c.
Postop mammography. Right MLO images. A spiculated density is seen on the right upper breast (a). Postop MRI. Axial and sagittal T2 weighted TSE, non-fat saturated image. A prominent susceptibility artefact is obscuring the right breast (b, c)
Discussion and Conclusion
Iron oxide technique is a new, promising safe and effective method used as an alternative to radiotracer method for sentinel lymph node detection (1–6). Large metanalyses such as Nordic SentiMag trial and The Central-European SentiMag study have shown their efficacy and safety (7–8). The Central-European SentiMag study found that more pathologically positive SLNs were found with the SentiMag technique compared to the radiotracer method (7). However, the artefact caused by iron oxide particles and its effect on the quality of follow up imaging have not been discussed in these trials. Although a couple of previous studies (11–13) have shown the susceptibility artefacts caused by iron oxide particles on MRI, the appearance of these particles on mammography which is the first line modality used during the follow ups after breast conserving surgery have not been previously discussed. To our knowledge, this is the first case study to demonstrate the iron oxide particles on mammography. Knowing the appearance of iron deposits on MRI and especially mammography is essential. In our case, the iron deposits were highly dense and denser than the breast parenchyma on mammography but interestingly occult on sonography. The density depends on the amount of particles accumulated; hence it might not be so dramatic in all cases. Iron oxide particles cause significant magnetic susceptibility as they are paramagnetic. On MRI, susceptibility artefact is seen as loss of signal which is called as “signal void” and they cause spatial distortion. The area effected by artefact is usually much larger than the size of the object causing the artefact so even a tiny amount of ferromagnetic material occult on x-ray or sonography can lower image quality on MRI. In the case study by Karakatsanis et al. (13), a patient with a history of iron oxide injection was discussed. In their case, unlike ours, no artefact was seen on the follow up mammogram. However, susceptibility artefacts were observed on the follow up MRI. On post contrast MR images, a new tumour was seen adjacent to artefacts. In our case the artefact on MRI was so evident in all sequences that it was impossible to detect a new lesion. Besides artefacts, staining the skin on the injected area (14) which is usually temporary is another limitation of these particles. Interestingly no staining was observed in our case.
We believe lowering the doses of these particles might be a solution for these undesirable side effects. An ongoing trial (Senti-Dose, https://doi.org/10.1186/ISRCTN11156955) is currently investigating the effect of lower doses.
In conclusion, the artefacts caused by iron deposits might be a problem in the long run. Reducing the amount of the injected material or changing the technique of massage for dissolving the particles might be a solution. As radiologists, we have to be alert for these kind of side artefacts. When we deal with a new density on follow ups of breast cancer on mammography, we have to question the method used for SLNB to avoid unnecessary biopsies and more importantly, to reduce the patient’s anxiety. Knowing the patient’s history and the method used for SLN detection guide us to consider this differential diagnosis.
Footnotes
Informed Consent: Written informed consent was obtained from patient who participated in this study.
Peer-review: Externally peer-reviewed.
Author Contributions: Concept - L.Ç.; Design - G.A.; Supervision - L.Ç.; Data Collection and/or Processing - G.A.; Analysis and/or Interpretation - L.Ç., GA.; Literature Search - G.A.; Writing Manuscript - G.A.; Critical Review - L.Ç.
Conflict of Interest: The authors have no conflicts of interest to declare.
Financial Disclosure: The authors declared that this study has received no financial support.
References
- 1.Veronesi U, Paganelli G, Viale G, Luini A, Zurrida S, Galimberti V, Intra M, Veronesi P, Robertson C, Maisonneuve P, Renne G, De Cicco C, De Lucia F, Gennari R. A randomized comparison of sentinel-node biopsy with routine axillary dissection in breast cancer. N Engl J Med. 2003;349:546–553. doi: 10.1056/NEJMoa012782. [DOI] [PubMed] [Google Scholar]
- 2.Mansel RE, Fallowfield L, Kissin M, Goyal A, Newcombe RG, Dixon JM, Yiangou C, Horgan K, Bundred N, Monypenny I, England D, Sibbering M, Abdullah TI, Barr L, Chetty U, Sinnett DH, Fleissig A, Clarke D, Ell PJ. Randomized multicenter trial of sentinel node biopsy versus standard axillary treatment in operable breast cancer: the ALMANAC Trial. J Natl Cancer Inst. 2006;98:599–609. doi: 10.1093/jnci/djj158. [DOI] [PubMed] [Google Scholar]
- 3.Zavagno G, De Salvo GL, Scalco G, Bozza F, Barutta L, Del Bianco P, Renier M, Racano C, Carraro P, Nitti D. GIVOM Trialists. A Randomized clinical trial on sentinel lymph node biopsy versus axillary lymph node dissection in breast cancer: results of the Sentinella/GIVOM trial. Ann Surg. 2008;247:207–213. doi: 10.1097/SLA.0b013e31812e6a73. [DOI] [PubMed] [Google Scholar]
- 4.Gill G SNAC Trial Group of the Royal Australasian College of Surgeons (RACS), NHMRC Clinical Trials Centre. Sentinel-lymph-node-based management or routine axillary clearance? One-year outcomes of sentinel node biopsy versus axillary clearance (SNAC): a randomized controlled surgical trial. Ann Surg Oncol. 2009;16:266–275. doi: 10.1245/s10434-008-0229-z. [DOI] [PubMed] [Google Scholar]
- 5.Krag DN, Anderson SJ, Julian TB, Brown AM, Harlow SP, Ashikaga T, Weaver DL, Miller BJ, Jalovec LM, Frazier TG, Noyes RD, Robidoux A, Scarth HM, Mammolito DM, McCready DR, Mamounas EP, Costantino JP, Wolmark N National Surgical Adjuvant Breast and Bowel Project. Technical outcomes of sentinel-lymph-node resection and conventional axillary-lymph-node dissection in patients with clinically node-negative breast cancer: results from the NSABP B-32 randomised phase III trial. Lancet Oncol. 2007;8:881–888. doi: 10.1016/S1470-2045(07)70278-4. [DOI] [PubMed] [Google Scholar]
- 6.Canavese G, Catturich A, Vecchio C, Tomei D, Gipponi M, Villa G, Carli F, Bruzzi P, Dozin B. Sentinel node biopsy compared with complete axillary dissection for staging early breast cancer with clinically negative lymph nodes: results of randomized trial. Ann Oncol. 2009;20:1001–1007. doi: 10.1093/annonc/mdn746. [DOI] [PubMed] [Google Scholar]
- 7.Thill M, Kurylcio A, Welter R, van Haasteren V, Grosse B, Berclaz G, Polkowski W, Hauser N. The Central-European SentiMag study: sentinel lymph node biopsy with superparamagnetic iron oxide (SPIO) vs radioisotope. Breast. 2014;23:175–179. doi: 10.1016/j.breast.2014.01.004. [DOI] [PubMed] [Google Scholar]
- 8.Karakatsanis A, Christiansen PM, Fischer L, Hedin C, Pistioli L, Sund M, Rasmussen NR, Jørnsgård H, Tegnelius D, Eriksson S, Daskalakis K, Wärnberg F, Markopoulos CJ, Bergkvist L. The Nordic SentiMag trial: a comparison of super paramagnetic iron oxide (SPIO) nanoparticles versus Tc99 and patent blue in the detection of sentinel node (SN) in patients with breast cancer and a meta-analysis of earlier studies. Breast Cancer Res Treat. 2016;157:281–294. doi: 10.1007/s10549-016-3809-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Teshome M, Wei C, Hunt KK, Thompson A, Rodriguez K, Mittendorf EA. Use of a Magnetic Tracer for Sentinel Lymph Node Detection in Early-Stage Breast Cancer Patients: A Meta-analysis. Ann Surg Oncol. 2016;23:1508–1514. doi: 10.1245/s10434-016-5135-1. [DOI] [PubMed] [Google Scholar]
- 10.Ghilli M, Carretta E, Di Filippo F, Battaglia C, Fustaino L, Galanou I, Di Filippo S, Rucci P, Fantini MP, Roncella M. The superparamagnetic iron oxide tracer: a valid alternative in sentinel node biopsy for breast cancer treatment. Eur J Cancer Care. 2017:26. doi: 10.1111/ecc.12385. [DOI] [PubMed] [Google Scholar]
- 11.Huizing E, Anninga B, Young P, Monypenny I, Hall-Craggs M, Douek M. Analysis of void artefacts in post-operative breast MRI due to residual SPIO after magnetic SLNB in SentiMAG Trial participants. Eur J Surg Oncol. 2015;41:S18. doi: 10.1016/j.ejso.2015.03.005. [DOI] [Google Scholar]
- 12.Krischer B, Forte S, Niemann T, Kubik-Huch RA, Leo C. Feasibility of breast MRI after sentinel procedure for breast cancer with superparamagnetic tracers. Eur J Surg Oncol. 2018;44:74–79. doi: 10.1016/j.ejso.2017.11.016. [DOI] [PubMed] [Google Scholar]
- 13.Karakatsanis A, Obondo C, Abdsaleh S, Hersi AF, Eriksson S, Wärnberg F. Optimisation of breast MRI compatibility after sentinel node biopsy with paramagnetic tracers. Eur J Surg Oncol. 2018;44:731–732. doi: 10.1016/j.ejso.2018.01.241. [DOI] [PubMed] [Google Scholar]
- 14.Karakatsanis A, Daskalakis K, Stålberg P, Olofsson H, Andersson Y, Eriksson S, Bergkvist L, Wärnberg F. Superparamagnetic iron oxide nanoparticles as the sole method for sentinel node biopsy detection in patients with breast cancer. Br J Surg. 2017;104:1675–1685. doi: 10.1002/bjs.1060. [DOI] [PubMed] [Google Scholar]