cyanotoxin recognition element assay format electrode novelty/biosensor improvement electroch. technique working range (ng/mL) LOD (pg/mL) water samples ref
MC-LR mAb direct chitosan modified Support; PAMAM, Ag NCs DPV (0.05–25) × 103 17 LW (173)
MC-LR PAb direct gold Support; Au NPs EIS 0.05–300 18.2 DI (174)
MC-LR PAb direct GCE Support; graphene-Au NP nanocomposite and polyDPB, and IL DPV (1 × 10−7)–(8 × 10−8) 0.000037 TW, LW, RW, GW (45)
MC-LR mAb direct IDE Support; SWCNTs chemiresistance 0.001–1 0.6 TW LW, RW (87)
MC-LR PAb direct GCE Support; MWCNT/IL EIS 0.005–1 1.7 DI, BW, TW (175)
MC-LR Mab direct Au Support; Au NPs DPV 0.05–15 20 Crude algae (68)
MC-LR mAb indirect competitive MWCNTs MWCNTs probe EIS 0.05–20 40 DI (176)
MC-LR mAb indirect competitive 3D foam G Graphene probe EIS 0.05–20 50 TW (177)
MC-LR mAb direct Au Flow-through, support; thiourea SAM/Ag NPs potentiometric step (capacitance) 10−5–1 0.007 BW, TW, Raw W (46)
SXT mAb noncompetitive Cu-graphene Flow-through, support; polymeric lipid membrane/graphene potentiometric 1–1000 1000 LW, shellfish (55)