Table 1.

CD4 count bias between FTCA and reference methods for some clinical samples with high and low monocyte concentrations.

Sample No. Sample type Monocytes count, cells/μL CD4 count, cells/μL Bias (%)
Reference methods FTCA*
1 CD3 depleted 800 220 205 −15 (−6.8 %)
2 CD3 depleted 800 204 188 −16 (−7.8 %)
3 Diluted blood 160 210 194 −16 (−7.6 %)
4 Whole blood 730 451 477 +26 (+5.8 %)
5 Whole blood 440 409 391 −18 (−4.4 %)
*

Calculated from formula: y = 0.336 x + 56.2 (from calibration plot Trendline)