. 2019 Jul;21(4):658–676. doi: 10.1016/j.jmoldx.2019.03.006

Table 2.

Extraction and Quantitation Methods and DNA Yields for the Reference Material Samples

Laboratory Extraction kit Elution buffer Quantitation method DNA yield per VAF, ng/mL
0% 0.125% 0.50% 2%
A Nucleosnap DNA Plasma Kit 5 mmol/L Tris, pH 8.5 Qubit HS kit 31.33 ± 0.12 23.73 ± 1.03 30.80 ± 2.96 24.07 ± 1.47
B QIAamp CNA Kit Kit supplied buffer Qubit HS kit 30.96 ± 4.15 14.60 ± 6.03 27.79 ± 8.21 21.17 ± 7.11
C Maxwell RSC LV ccfDNA Kit (automated) 10 mmol/L Tris, pH 8.0 Qubit HS kit 15.24 ± 2.27 12.01 ± 1.72 16.25 ± 2.18 9.18 ± 5.32
D QIAsymphony DSP Circulating DNA Kit (automated) Kit supplied buffer Qubit HS kit 18.20 ± 0.47 18.27 ± 5.85 17.53 ± 4.24 15.93 ± 0.28
E QIAamp CNA Kit 10 mmol/L Tris, pH8.0 Qubit HS kit 17.51± 5.19 11.45 ± 1.85 16.64 ± 1.66 11.75 ± 1.27
F QIAamp CNA Kit Kit supplied buffer Qubit BR kit 30.71 ± 4.32 26.78 ± 1.75 30.09 ± 4.21 27.31 ± 5.00
G Zymo Quick-cfDNA Serum and Plasma kit Water Qubit HS kit 4.34 ± 0.44 3.16 ±.94 4.32 ± 2.75 4.05 ± 0.34

Data are expressed as means ± SD. Samples were prepared in triplicate for the laboratories with the exception of laboratory G, where four samples were measured.

BR, broad range; ccfDNA, circulating cell-free DNA; cfDNA, cell-free DNA; CNA, circulating nucleic acid; DSP, digital spatial profiling; HS, high sensitivity.

Carrier RNA was added in the lysis step.