. 2019 Jul;21(4):718–733. doi: 10.1016/j.jmoldx.2019.03.002

Table 2.

Single-Site and Multisite Precision

Target MR Mean MR SD Target %IS Mean %IS % CV N
Single-site precision
 1 0.70 0.08 10 20.3613 16.6 40
 2 1.63 0.08 1 2.3423 11.7 40
 3 2.66 0.08 0.1 0.2211 11.0 40
 3.5 3.18 0.10 0.032 0.0663 16.8 40
 4 3.68 0.13 0.01 0.0215 18.9 40
Multisite precision
 1 0.74 0.06 10 18.1906 13.7 240
 2 1.69 0.07 1 2.0787 15.8 240
 3 2.70 0.09 0.1 0.2031 20.9 240
 3.5 3.22 0.13 0.032 0.0628 29.4 240
 4 3.72 0.13 0.01 0.0197 29.9 240

The precision of the assay was evaluated with both single-site and multisite studies by testing five separate dilution series across five target MR values. Data herein were combined per target MR level. SD values of MR measurements formed the definitive acceptance criteria. %IS values are shown for historical reference. Supplemental Tables S1 and S2 detail each specimen and source of variation separately.

IS, international scale; MR, molecular response.