Table 4.

LOQ Data

Specimen no. Transcript identity Valid measurements MR (mean) SD Result
1 e13a2 20 4.79 0.27 Pass
2 e14a2 20 4.67 0.23 Pass
3 e14a2 19 4.80 0.34 Pass
4 e14a2 20 4.60 0.24 Pass
5 e14a2 20 4.87 0.25 Pass
6 e13a2 20 4.82 0.25 Pass

The LOQ of the assay was validated as MR4.7 (0.002% international scale) by measuring six low-level specimens across 119 valid measurements using two lots of the kit across 3 testing days. All specimens satisfied the error criteria (SDMR ≤ 0.36), supporting a limit beyond (and, therefore, constrained by) the limit of detection of MR4.70.

LOQ, limit of quantitation; MR, molecular response.