Abstract
Cyanobacterial blooms cause local and global health issues by contaminating surface waters. Microcystins and nodularins are cyclic cyanobacterial peptide toxins comprising numerous natural variants. Most of them are potent hepatotoxins, tumor promoters, and at least microcystin-LR is possibly carcinogenic. In drinking water, the World Health Organization (WHO) recommended the provisional guideline value of 1 µg/L for microcystin-LR. For water used for recreational activity, the guidance values for microcystin concentration varies mostly between 4–25 µg/L in different countries. Current immunoassays or lateral flow strips for microcystin/nodularin are based on indirect competitive method, which are generally more prone to sample interference and sometimes hard to interpret compared to two-site immunoassays. Simple, sensitive, and easy to interpret user-friendly methods for first line screening of microcystin/nodularin near water sources are needed for assessment of water quality and safety. We describe the development of a two-site sandwich format lateral-flow assay for the rapid detection of microcystins and nodularin-R. A unique antibody fragment capable of broadly recognizing immunocomplexes consisting of a capture antibody bound to microcystins/nodularin-R was used to develop the simple lateral flow immunoassay. The assay can visually detect the major hepatotoxins (microcystin-LR, -dmLR, -RR, -dmRR, -YR, -LY, -LF -LW, and nodularin-R) at and below the concentration of 4 µg/L. The signal is directly proportional to the concentration of the respective toxin, and the use of alkaline phosphatase activity offers a cost efficient alternative by eliminating the need of toxin conjugates or other labeling system. The easy to interpret assay has the potential to serve as a microcystins/nodularin screening tool for those involved in water quality monitoring such as municipal authorities, researchers, as well as general public concerned of bathing water quality.
Keywords: noncompetitive immunoassay, cyanotoxin, microcystin, nodularin, lateral flow
1. Introduction
Blooms of cyanobacteria, a phylum of bacteria that obtain their energy through photosynthesis, are commonly found around the globe. Certain cyanobacterial strains are known to produce toxic secondary metabolites called cyanotoxins, which cause diverse problems by contaminating water sources used for recreation, drinking water production and (aqua) farming. Generally, the most widespread and problematic cyanotoxin group consists of cyclic peptide toxins comprising numerous natural variants. Close to 250 analogues of microcystin and about 10 analogues of nodularin with differing toxicity have been reported to date in the literature [1]. The heptapeptide, microcystin in fresh water, and pentapeptide nodularin in brackish water bodies are the potent hepatotoxins acting as specific inhibitors of protein phosphatases (PPs) and thus hazardous to health, being responsible for growth of tumor precursors, particularly in liver [2,3]. Cyanotoxins have the potential to accumulate in fish and marine organisms via food chains [4]. Microcystin and nodularin are very stable and resistant to hydrolysis, making them a persistent issue for common water usage [5]. The robustness results from the chemical structure of microcystin and nodularin, which consists of a monocyclic peptide ring and a hydrophobic Adda moiety (3-amino-9-methoxy-2,6,8-trimethyl-10-phenyldeca-4(E),6(E)-dienoic acid). The conserved Adda group is the common determinant for all microcystin and nodularin, and has proven useful in the development of broad-specificity detection methods [6,7]. Microcystin-LR is the widely distributed and most studied microcystin analogue. Similarly, nodularin-R is the most common nodularin variant found in the brackish water bodies such as in the Baltic sea. World Health Organization (WHO) has established guidelines for cyanotoxins in the assessment of drinking water quality and recreational use [8,9]. For drinking water, WHO set guideline value is 1 µg/L of microcystin-LR. In addition, several countries have issued guideline values for recreational water use where the microcystin-LR levels usually range between 4.0–25 µg/L [10].
The large number of toxin variants combined with the low-molecular weight of microcystins and nodularins (~1000 Da) creates major challenges for the development of efficient analytical detection methods. The confirmatory detection and identification of cyanotoxins is routinely done in centralized laboratories with the use of high-sensitivity reference methods such as liquid chromatography (LC) [11,12], mass-spectrometry (MS) [13], or a combination of both [14,15]. Although these methods are well suited for the exact identification of the contaminating toxin, the need for specialized instrumentation and trained staff to operate them makes them unsuitable for rapid or high-throughput testing. Alternative methods for cyanotoxin testing have also been established, particularly methods based on protein phosphatase inhibition (PPI) [16,17] and antibodies [6,18] are widely used due to the simplicity and cost-efficiency of the assays. Although these methods are widely used in commercial kits, both techniques have certain limitations. Usually, these methods are performed in laboratory and are not suited easily for on-site use.
One of the main benefits of using antibodies for cyanotoxin detection is the possibility to implement the assay format for field testing, as the result of the possible contamination is often needed rapidly, and the guideline values do not require exceptional sensitivity. Among such methods, lateral-flow immunoassays (LFIA) have gained considerable interest in the academic and industrial research [19,20]. LFIA immunostrips are widely used to target various biomolecules in a myriad of commercial applications, ranging from monitoring of drug abuse to environmental contaminants [21]. Lateral-flow assays provide rapid results in a very cost effective and de-centralized manner. The results for LFIA can be observed without the need of any special expertise and thus it offers wide applicability for testing location. Field testing is particularly useful for cyanotoxin detection, as easy monitoring tools for surface waters at multiple sites are needed to meet the established guideline values. However, the simultaneous and simple detection of all the different variants is extremely challenging and the low-molecular weight of cyanotoxins has limited the available detection methods used in LFIA to competitive assay formats [22,23]. Furthermore, competitive immunoassays are intrinsically more prone to sample interference and the results are hard to interpret compared to the noncompetitive counterparts. In addition, most of the currently available LFIA assays utilize conjugated toxin derivatives as a reporter system for the assay, which are often expensive to manufacture.
We describe the development of a simple chromogenic LFIA for simultaneous detection of microcystins and nodularin. The assay utilizes recombinant single-chain antibody fragment-alkaline phosphatase (scFv-AP) fusion protein as a reporter, which provides a readymade tracer system where the signal intensity on the immunostrip test line is directly proportional to the toxin concentration. Moreover, the result can be visually confirmed without the need for separate reading device. The sensitivity of the LFIA was found to be well below the 4 µg/L of microcystin-LR equivalents, the minimum guideline value set for recreational water use by certain countries such as Hungary. The LFIA was capable of simultaneous detection of microcystins and nodularin-R from environmental water samples collected at sites known to be contaminated with cyanotoxins during the cyanobacterial blooming season. The assay can be used as a screening tool for the detection of microcystin and nodularin from surface waters.
2. Materials and Methods
2.1. Common Materials and Reagents
Common inorganic and organic chemical reagents were obtained from commercial source Sigma or Merck unless otherwise specified. The reagent water used was purified by Millipore Milli-Q Plus water filtration purification system (Millipore Corporation, Burlington, MA, USA). Enhancement solution, wash concentrate, and streptavidin coated microtiter plates were from Kaivogen (Turku, Finland). Monoclonal anti-Adda antibody, AD4G2 (Adda specific, anti-Microcystins) was from Enzo Life Sciences, Inc. (Farmingdale, NY, USA). Bacterial anti alkaline phosphatase polyclonal antibody (anti-AP Pab) which was purchased from LifeSpan Biosciences, Inc. (Seattle, WA, USA) was purified in house through protein G affinity purification. Histidin tag scFv purification was done by His Spin Trap™ kit (GE Healthcare, Amersham, UK). The bacterial host Escherichia coli XL-1 Blue was from Stratagene, La Jolla, CA. Fc specific monoclonal human anti-mouse IgG (HAMA) which recognize mouse IgG via the Fc region was a gift from Dr. Keith Thompson (University of Oslo). Alkaline phosphatase substrate BCIP/NBT (5-Bromo-4-chloro-3-indolyl phosphate BCIP and nitro blue tetrazolium NBT) tablets were purchased from SIGMA. According to the manufacturer’s instruction, one tablet was dissolved in 10 mL of water yielding substrate solution of BCIP (0.15 mg/mL), NBT (0.30 mg/mL), Tris buffer (100 mM), and MgCl2 (5 mM), pH 9.25–9.75. Lateral flow assay buffer (LFAB) was composed of 10 mM Phosphate, 137 mM NaCl, 2.7 mM KCl, pH 7.3; supplemented with 0.5% Tween-20, 1% BSA, 0.06% bovine γ-albumin, and filtered through a 0.22 µm filter. Once prepared, it was kept at 4 °C and used for two weeks. Three times LFAB (3 × LFAB) was prepared using the above composition with three times molar excess. Superb broth (SB medium, pH 7) was composed of 2% yeast extract, 3% tryptone, and 1% MOPS.
2.2. Instrumentation
Multilabel counter VictorTM 1420 for fluorescence measurement was from PerkinElmer Life Sciences, Finland. Protein concentration were measured by NanoDrop ND1000 spectrophotometer (Thermo Scienctific, Waltham, MA, USA). A Linomat 5 sample applicator (CAMAG, Muttenz, Switzerland) was used for striping of the binder and control line molecule. A desktop paper cutter (Ideal 1058, Krug & Priester, Balingen, Germany) was used to cut the test strips.
2.3. Toxin Standards
Specific amount of the purified toxins were obtained from Dr. Jussi Meriluoto’s Lab (Åbo Akademi University) as a lyophilized dried powder (microcystin-LR, -dmLR, -RR, -dmRR, -YR, -LY, -LF -LW, nodularin-R, and anatoxin-a) or as solution (cylindrospermopsin). The microcystins and nodularins were purified as described previously [24]. Dry powder of microcystin and nodularin was dissolved in 50% methanol yielding 100 to 250 µM original stock. Dried anatoxin-a was dissolved in reagent water (~10 µg/mL original stock). Further working standard stocks of all toxins were diluted in reagent water and kept at −20 °C for long term or at 4 °C for short term in sealed condition.
2.4. Anti-Immunocomplex Antibody Fragment
The generic anti-immunocomplex (anti-IC) single-chain fragment (scFv) SA51D1 as fusion to alkaline phosphatase (scFv-AP SA51D1) reported in Akter et al., 2016 [25] was used in this work to develop the non-competitive sandwich-type LFIA. The isolation, purification and characterization of the anti-IC scFv-AP has been described in detail in Akter et al., 2016 [25]. The scFv-AP was expressed in XL-1 Blue E. coli cells in 50 mL culture in SB medium supplemented with 100 µg/mL ampicillin, 10 µg/mL tetracycline, 0.05% glucose, and induced at 26 °C for 4–6 h. Harvested cells were purified through histidin tagged scFv-AP using His trap affinity column (GE Healthcare) according to the manufacturer’s instructions. In Akter et al., 2016, [25] we reported the use of the anti-IC scFv-AP to develop a highly sensitive time-resolved fluoroscence based IC assay (TRF-IC assay) capable of detecting all the tested 11 different cyanobacterial peptide hepatotoxin (microcystin-LR, -dmLR, -RR, -dmRR, -WR, -YR, -LA, -LY, -LF, -LW, and nodularin-R) well below WHO guide line limit of 1 µg/L. The scFv-AP does not have any significant binding affinity towards naked anti-Adda Mab nor to the toxin alone [25]. Furthermore, using the scFv-AP, we also reported a non-competitive ELISA with broad specificity for microcystin and nodularin utilizing the fusion AP enzyme to generate visually detectable signal [26].
2.5. Preparation of Immunostrips
The lateral-flow immunostrip (Figure 1) used to perform the LFIA consists of cellulose absorption pad (Millipore, USA), nitrocellulose membrane (Hi-Flow Plus HF180, Millipore, USA), and glass fiber feeding pad (Millipore, USA).
A 2.5 cm wide nitrocellulose membrane was attached to the plastic adhesive backing card (G&L Precision Die Cutting, San Jose, CA, USA). An adsorption pad was prepared by attaching a 34 mm wide cellulose membrane strip (Millipore, MA, USA) to overlap 2–3 mm with the nitrocellulose membrane. A feeding pad of 16 mm wide glass fiber strip (Millipore, MA, USA) was attached to overlap 2–3 mm with the other end of the nitrocellulose membrane.
The striping/printing of the test line was performed by dispensing 0.5 mg/mL of HAMA in 10 mM Tris–HCl buffer (pH 8.0) with 1% methanol on the nitrocellulose membrane. The control line was printed similarly at a 5-mm distance by dispensing 0.6 mg/mL anti-AP Pab in the same buffer composition. Printing was accomplished using Linomat 5 sample applicator (CAMAG, Muttenz, Switzerland) which was adjusted to produce 1µL/cm stripes with a liquid flow speed of 250 nL/s. Membranes were then dried at 37 °C for 2–3 h. The assembled LFIA cards were then cut into 4 or 5 mm wide chips/strips using a desktop paper cutter (Ideal 1058, Krug & Priester, Balingen, Germany). In each experiment, same-width strips were used.
2.6. Noncompetitive Sandwich-Type LFIA Procedure
The LFIA was performed in a standard 96 well polypropylene microtiter plate (Greiner BioOne, Kremsmünster, Austria) and consisted of four steps (Figure 1b). (1) A pre-incubation step: 40 µL of reagent water (as blank)/sample/toxin standard was added to a 20 µL of reaction mix (20 ng anti-Adda Mab and 35 ng anti-IC scFv-AP SA51D01 in 3 × LFAB) and incubated 10–15 min at slow shaking. (2) Feeding step: In this step the feeding pad of the immunostrips were dipped in the total 60 µL reaction volume. The absorption of the liquid took place within approximately 10 min. (3) Washing step: In the washing step, stripes were moved into a well containing 100 µL of LFAB and left for 25 min. (4) Colorimetric reaction: In this fourth step, the strips were moved to the reaction well containing 150 µL of BCIP/NBT substrate solution. The absorption and colorimetric reaction was completed within about 2 h. All steps were carried out in room temperature (RT). The detailed principle and procedure of the chromogenic LFIA is described in Figure 1.
2.7. Applicability of the LFIA in the Detection of Different Hepatotoxins
Nine different common cyanobacterial hepatotoxins (heptapeptide microcystin-LR, -dmLR, -RR, -dmRR, -YR, -LY, -LF, -LW, and pentapeptide nodularin-R) were used to assess the specificity of the LFIA. The toxins were spiked in to LFAB at the concentration of 4 µg/L. The LFIA procedure was carried out as described earlier in Section 2.6.
2.8. Detection of Microcystin-LR from Spiked Water Samples
Reagent water and two raw environmental surface water samples from Finnish lake (S11: Alasenjärvi, Lahti, 6 July, 2009 and V1: Rusutjärvi, 23 June, 2009) were spiked with microcystin-LR over a range of concentrations (0, 1, 4, 10, 20 µg/L). The used environmental samples collected in 2009 were previously analyzed for intracellular toxin content by LC-MS and were stored at −20 °C. Trace amount of cyanobacterial hepatotoxin (0.07 and 0.04 µg/L, respectively) was detected from these two samples. The LFIA was performed using duplicate strips according to procedure described in Section 2.6.
2.9. LFIA Performance with Environmental Water Samples
Fourteen raw surface water samples collected during 2009 from Finland and Estonia were analyzed by the LFIA to detect total toxin (extracellular and intracellular). The samples were frozen and thawed at least twice to break the cells and to allow the toxins to be released in the samples. No further extraction or process was performed on the samples. Furthermore, the same raw water samples were analyzed according to the previously reported TRF-IC assay [26]. In addition, for each corresponding samples, intracellular toxin concentration result (using extracted cell samples) by LC-MS was available which was reported earlier by Hautala et al. [27].
2.10. LFIA Performance with Possible Interference
Reagent water was spiked with microcystin-LR, anatoxin-a and cylindrospermopsin at concentration ranging 1 to 100 µg/L of toxin. The LFIA was performed using duplicate strips according to procedure described in Section 2.6. Water sample containing no toxin was tested as a control.
2.11. Data Analysis and Result Interpretation
The results of the LFIA was done by visual inspection, and the images were recorded using Canon EOS 60D DSLR (Canon Inc., Tokyo, Japan), Samsung Galaxy A5 2016 mobile phone camera (Samsung, Seoul, South Korea) and ChemiDoc MP Imaging System (Bio-Rad, Hercules, CA, USA). The images were processed with Corel Photo-Paint 2018 software (Corel Corporation, Ottawa, ON, Canada). Briefly, the images were cropped and adjusted parallel for illustrative purposes and the images were converted to greyscale.
3. Results
3.1. LFIA Development and Detection of Microcystin-LR from Spiked Samples
We spiked three different water sources (reagent water and two environmental water samples) with known amount of microcystin-LR to assess the initial performance of the LFIA immunostrips. No significant difference could be observed among the water sources or the replicate strips done (Figure 2). For the test, microcystin-LR was selected due to its wide distribution in environmental samples. All of the four microcystin-LR concentrations used in the test resulted in a clear positive test line (Figure 2). Signal intensity increased proportionally with increasing toxin concentration. In addition, the control line could be detected in all immunostrips.
3.2. Applicability of the LFIA in the Detection of Different Hepatotoxins
The specificity of the anti-IC scFv-AP SA51D1 towards different cyanobacterial microcystins was extremely board and as reported previously in Akter et al., the scFv-AP was capable of detecting all of the tested 11 different cyanotoxins (microcystin-LR, -dmLR, -RR, -dmRR, -WR, -YR, -LA, -LY, -LF, -LW, and nodularin-R) [25]. The board specificity profile was confirmed also for the LFIA with the use of nine different cyanotoxins (nodularin-R; and microcystin-LR, -dmLR, -RR, -dmRR, -LF, -LY, -LW, and -YR) as analytes. The LFIA strips (see Supplementary Materials) were used to analyze the 4 µg/L of toxin analyte. All of the analyzed samples gave clear positive test lines (Figure 3).
3.3. LFIA Performance with Enviromental Samples
The LFIA performance with natural samples was tested with the use of 14 environmental samples collected from various locations during the blooming season in southwest Finland and Estonia. For each sampling site, two parallel type of samples were available: untreated raw water and filtered cell. The raw waters were analyzed with the developed LFIA as well as with TRF-IC assay [25]. Intracellular toxin content from extracted cells were previously analyzed by LC-MS and was reported by Hautala et al. [27]. The details of the samples are summarized in Table 1.
Table 1.
Sample | Sample Location | Date | TRF-IC Assay 1 (µg/L) | LC-MS 2 (µg/L) Adapted from Hautala et al. [27] |
---|---|---|---|---|
toxin content from raw water | intracellular toxin content from extracted cells | |||
Å15 | Vandö kanal, Finström, Åland Islands, Finland | 28.07.2009 | 0 | 0 |
Å21 | Nåtö hemviken, Nåtö Island, Åland Islands, Finland | 30.07.2009 | 10.1 | 8.6 |
Å22 | Nåtö vägbank (sea), Åland Islands, Finland | 29.07.2009 | 0.9 | 1.5 |
Å23 | Dalkarby träsk, Dalkarby, Åland Islands, Finland | 29.07.2009 | 0 | 0 |
L6 | Littoistenjärvi, Kaarina, Finland | 26.08.2009 | 3.6 | 5.2 |
L8 | Littoistenjärvi, Kaarina, Finland | 11.09.2009 | 4.2 | 3.7 |
E17 | Lake Harku, Estonia | 18.08.2009 | 4.2 | 1.97 |
RN13 | Hauninen reservoir, Raisio, Finland | 9.06.2009 | 5.2 | 11.9 |
RN14 | Hauninen reservoir, Raisio, Finland | 16.06.2009 | 9.4 | 23.6 |
RN15 | Hauninen reservoir, Raisio, Finland | 23.09.2009 | 9.1 | 21.7 |
RN26 | Hauninen reservoir, Raisio, Finland | 1.09.2009 | 0.7 | 0.83 |
RN30 | Hauninen reservoir, Raisio, Finland | 29.09.2009 | 1.8 | 1.9 |
T3 | Savojärvi, Pöytyä, Finland | 7.08.2009 | 28.4 | 40.9 |
T6 | Maaria reservoir, Turku, Finland | 24.08.2009 | 39.8 | 49.4 |
The LFIA immunostrips results were visually inspected (Figure 4). Two of the 14 samples previously confirmed as negative for all cyanotoxin variants [27] yielded no visible test lines. The positive samples based on reference method yielded faint to intense test lines in LFIA immunostrips depending on the toxin concentration. Two of the sampling locations contained heavy blooming of algae and consequentially had high concentrations of cyanotoxins (>25 µg/L).
3.4. LFIA Performance with Possible Interference
The LFIA performance with possible interfering agents are investigated with the use of other cyanobacterial toxin such as anatoxin-a and cylindrospermopsin. As expected, the presence of high amount (100 µg/L) of anatoxin-a and cylindrospermopsin did not result in any false positive or inconclusive decision (Figure 5).
4. Discussion
The adverse effects caused by freshwater cyanobacteria are of major concern as some of the metabolites are toxic to humans and other mammals, hamper plant growth, and may accumulate in sea food. The diversity and complexity of the toxins produced by cyanobacteria creates challenges, especially for field testing. Although the reference methods, such as MS (mass spectrometry) and HPLC (high-performance liquid chromatography) for the detection of microcystin and nodularin are well established, more cost-efficient, and rapid methods are urgently needed for field testing. Our goal was to develop a broadly specific noncompetitive LFIA for the simultaneous detection of microcystin and nodularin directly from untreated environmental samples.
The lateral-flow immunostrips were designed to provide a simple and reproducible result with the use of recombinant antibody fragment in fusion with bacterial AP. The use of AP as a reporter on the immunostrip had two clear benefits over previously described LFIA systems. (1) The scFv-AP as a chromogenic reporter in the LFIA enabled the simple design of the immunostrip, where the readymade tracer can be produced with simple bacterial expression. This removes the need for any chemical conjugation steps of additional reporter molecules, and at the same time, keeping the molar ratio of the antibody-label constant. (2) The excess scFv-AP not bound to the immunocomplex of the anti-Adda Mab and the cyanotoxin was used to establish the control line for the assay. The use of HAMA instead of the Adda Mab on the test line was done to ensure the formation of the IC before the lateral-flow step. Since the formation of the IC is dependent on the affinities of the parental Adda Mab to the cyanotoxin, and the scFv, to the IC, the preincubation time was used to gain higher sensitivity in the system. The substrate BCIP/NBT has previously been successfully implemented for paper based immunoassays [28,29]. Different AP substrates were tested for the LFIA during this study (result not shown), and the benefits of BCIP/NBT were confirmed to be superior in comparison to other substrates. As the reaction end product precipitates due to the AP activity, it restricts the diffusion of the signal outside the detection lines. In addition, the reaction end product has a very intense blue-purple color, which forms a good contrast against white nitrocellulose LF-matrix. Moreover, the product is stable and not light sensitive, making it suitable also for field testing.
The immunocomplex formed by the anti-Adda Mab and the toxin was detected with the use of previously characterized antibody-fragment, which is known to bind the formed immunocomplex with uniquely broad specificity for microcystin and nodularin. The possibility for a simultaneous detection of multiple cyanotoxins from different toxin class in a noncompetitive format is ideal for high-throughput testing of samples where the contaminating cyanotoxin is unknown. Our aim was to develop this broad-specificity assay concept in to even simpler format, where the result could be obtained rapidly without the need for any specialized instrumentation. The amounts of anti-Adda Mab (20 ng) and the generic anti-IC scFv-AP (35 ng) were optimized to generate strong enough signals for visual inspection on the test and control lines. Next, we tested the performance of the LFIA by spiking variable amounts of microcystin-LR toxin to reagent water as well as environmental waters. As shown in Figure 2, the increasing amount of microcystin-LR correlated strongly with the signal intensity observed on the test line. This contrasts with currently available lateral flow tests for hepatotoxins, where the test line slowly fades with increasing toxin concentration. Although the signal intensity decreases on the test line with lower microcystin-LR concentrations (<1 µg/L), the positive result can still be confirmed from samples containing as low as 0.7 µg/L of cyanotoxins based on the reference methods (LC-MS).
The previously reported TRF-IC immunoassays [25] have shown that, in addition to the high sensitivity of the assay, it also performs with a very broad specificity to hepatotoxins containing the Adda moiety. We wanted to confirm this result also with the use of LFIA by testing the cross-reactivity of the assay using eight different microcystin congeners, nodularin-R, and a mixture of the all nine different hepatotoxins. The concentration of each toxin was set to 4 µg/L, and all the tested toxin samples contributed clear positive results (Figure 3). This confirms the broad-specificity performance of the IC-assay concept, also with simpler assay concept.
The final step for the LFIA development was to test the performance of the immunostrips with the use of environmental samples. The use of real samples where the toxin concentration is unknown can create problems for the accurate detection of cyanotoxins mainly resulting from cross-reacting compounds causing interferences and unspecific matrix effects. In addition, some of the cyanobacterial strains can produce multiple classes of toxins simultaneously. We tested the LFIA in the presence of anatoxin-a and cylindrospermopsin at concentration range from 1–100 µg/L and no interference is observed (Figure 5). We used 14 samples collected from various locations in Finland and Estonia during the cyanobacterial blooming season (June–September). The sampling locations consisted of both fresh and brackish water areas where cyanotoxins are known to be present. The raw waters used for the LFIA were also analyzed by the previously reported TRF-IC assay [25]. Furthermore, all of the samples were previously analyzed [27] with LC-MS methods for the intracellular toxin content, and the detailed description of sampling locations are described in Table 1. All the samples containing cyanotoxin levels > 4µg/L produced a clear positive test line on the immunostrips (Figure 4). For samples containing toxin concentrations ranging between 0.7–4 µg/L depending on the analytical approach (Å22, RN26, and RN30), the bands are faint but observable. No false positive or false negative was observed by the LFIA. Overall, the test line results for LFIA immunostrips correlate well with the previously found cyanotoxin concentration obtained with confirmatory methods. In future, the LFIA described in this study can be further optimized and implemented also for the detection of cyanobacterial contamination of drinking water, which is a major concern in countries with inadequate water treatment facilities.
5. Conclusions
An easy to interpret noncompetitive LFIA test for microcystins and nodularin-R was successfully developed. The assay can detect wide variety of cyanobacterial hepatotoxin well below 4 µg/L which is the lowest detection value used for recreational water in many countries. The LFIA is suitable for field conditions and the result can be visually confirmed with no need for a separate measuring device for result interpretation. The signal intensity is directly proportional to the concentration of toxin. Based on conventional AP activity, the assay offers a cost-effective method for first line screening of microcystins and nodularin for assessment of water quality and safety.
Acknowledgments
We thank Jussi Meriluoto for providing cyanotoxin reference materials. We thank Keith Thompson (University of Oslo) for proving us Fc specific monoclonal human anti-mouse IgG (HAMA) as a gift. We thank to those colleagues who provided samples: Sonja Nybom (Åbo Akademi University), Pirkko Ala-Uotila (Raisio-Naantali Waterworks), Pirkko Pajakko (Turku Municipal Water Company), Marko Järvinen (Finnish Environment Institute), Kirsti Lahti (Water Protection Association of the River Vantaa and Helsinki Region), and Aune Annus and coworkers (Tervisekaitse, Estonia).
Supplementary Materials
The following are available online at https://www.mdpi.com/2079-6374/9/2/79/s1. Figures S1–S3: Raw images of the LFIA strips. Figure S1. LFIA with blank and environmental samples set1; Figure S2. LFIA with environmental samples set2; Figure S3. LFIA with environmental samples set3.
Author Contributions
S.A., M.V., and U.L. were involved in conceptualization. S.A. and J.L. were involved in original draft preparation. S.A. and J.L. wrote and edited the manuscript. S.A. supervised and was involved in experimental planning. S.A. and T.K. designed and performed the experiments, collected data and picture. G.M. acquired his own funding and performed experiments. S.A., M.V., and U.L. were involved in funding acquisition. S.A. and M.V. were involved in project administration. T.K., G.M., and M.V. commented on the manuscript. S.A., J.L., and U.L. reviewed the manuscript. All gave consent on the manuscript.
Funding
This research work received financial support from TEKES, the National Technology Agency of Finland (grant 823/31/2014) and from Ministry of Agriculture and Forestry of Finland (Dnro ESAELY/721/2017). SA gratefully acknowledges research grants (24808 and 36427) from Maa-ja vesitekniikan tuki ry (MVTT). GM acknowledges Department of Science and Technology, Govt. of India, for DST-BOYSCAST overseas fellowship, (SR/BY/L-15/10).
Conflicts of Interest
S.A., M.V., and U.L. are inventors in a pending patent application WO2017109290A1 concerning the anti-immunocomplex antibody used to develop the lateral flow assay. University of Turku is the assignee of the application. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or in the decision to publish the results.
References
- 1.Meriluoto J., Spoof L., Codd G.A. Handbook of Cyanobacterial Monitoring and Cyanotoxin Analysis. John Wiley & Sons; Hoboken, NJ, USA: 2017. [Google Scholar]
- 2.Dawson R.M. The Toxicology of Microcystins. Elsevier; Amsterdam, The Netherlands: 1998. pp. 953–962. [DOI] [PubMed] [Google Scholar]
- 3.Stewart I., Seawright A.A., Shaw G.R. Anonymous Cyanobacterial Harmful Algal Blooms: State of the Science and Research Needs. Springer; Berlin, Germany: 2008. Cyanobacterial poisoning in livestock, wild mammals and birds–An overview; pp. 613–637. [DOI] [PubMed] [Google Scholar]
- 4.Ibelings B.W., Chorus I. Accumulation of cyanobacterial toxins in freshwater “seafood” and its consequences for public health: A review. Environ. Pollut. 2007;150:177–192. doi: 10.1016/j.envpol.2007.04.012. [DOI] [PubMed] [Google Scholar]
- 5.Chorus I., Bartram J. Toxic Cyanobacteria in Water: A Guide to Their Public Health Consequences, Monitoring and Management. CRC Press; Boca Raton, FL, USA: 1999. [Google Scholar]
- 6.Weller M.G., Zeck A., Eikenberg A., Nagata S., Ueno Y., Niessner R. Development of a direct competitive microcystin immunoassay of broad specificity. Anal. Sci. 2001;17:1445–1448. doi: 10.2116/analsci.17.1445. [DOI] [PubMed] [Google Scholar]
- 7.Fischer W.J., Garthwaite I., Miles C.O., Ross K.M., Aggen J.B., Chamberlin A.R., Towers N.R., Dietrich D.R. Congener-Independent immunoassay for microcystins and nodularins. Environ. Sci. Technol. 2001;35:4849–4856. doi: 10.1021/es011182f. [DOI] [PubMed] [Google Scholar]
- 8.World Health Organization . Guidelines for Drinking-Water Quality. World Health Organization; Geneva, Switzerland: 2004. [Google Scholar]
- 9.World Health Organization . Guidelines for Safe Recreational Water Environments: Coastal and Fresh Waters. World Health Organization; Geneva, Switzerland: 2003. [Google Scholar]
- 10.Ibelings B.W., Backer L.C., Kardinaal W.E.A., Chorus I. Current approaches to cyanotoxin risk assessment and risk management around the globe. Harmful Algae. 2014;49:63–74. doi: 10.1016/j.hal.2014.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Meriluoto J.A., Eriksson J.E. Rapid analysis of peptide toxins in cyanobacteria. J. Chromatogr. A. 1988;438:93–99. doi: 10.1016/S0021-9673(00)90236-X. [DOI] [PubMed] [Google Scholar]
- 12.Jones G.J. Release and degradation of microcystin following algicide treatment of a Microcystis aeruginosa bloom in a recreational lake, as determined by HPLC and protein phosphatase inhibition assay. Water Res. 1994;28:871–876. doi: 10.1016/0043-1354(94)90093-0. [DOI] [Google Scholar]
- 13.Erhard M., Von Döhren H., Jungblut P. Rapid typing and elucidation of new secondary metabolites of intact cyanobacteria using MALDI-TOF mass spectrometry. Nat. Biotechnol. 1997;15:906–909. doi: 10.1038/nbt0997-906. [DOI] [PubMed] [Google Scholar]
- 14.Jurczak T., Karlsson K., Meriluoto J. Characterization and diversity of cyano-bacterial hepatotoxins (Microcystins) in blooms from polish Freshwaters identified by liquid chromatography-electrospray ionisation mass spectrometry. Chromatographia. 2004;59:571–578. doi: 10.1365/s10337-004-0279-8. [DOI] [Google Scholar]
- 15.Meriluoto J., Karlsson K., Spoof L. High-throughput screening of ten microcystins and nodularins, cyanobacterial peptide hepatotoxins, by reversed-phase liquid chromatography-electrospray ionisation mass spectrometry. Chromatographia. 2004;59:291–298. [Google Scholar]
- 16.Mackintosh C., Beattie K.A., Klumpp S., Cohen P., Codd G.A. Cyanobacterial microcystin-LR is a potent and specific inhibitor of protein phosphatases 1 and 2A from both mammals and higher plants. FEBS Lett. 1990;264:187–192. doi: 10.1016/0014-5793(90)80245-E. [DOI] [PubMed] [Google Scholar]
- 17.Bouaïcha N., Maatouk I., Vincent G., Levi Y. A colorimetric and fluorometric microplate assay for the detection of microcystin-LR in drinking water without preconcentration. Food Chem. Toxicol. 2002;40:1677–1683. doi: 10.1016/S0278-6915(02)00103-5. [DOI] [PubMed] [Google Scholar]
- 18.Brooks W.P., Codd G.A. Immunoassay of hepatotoxic cultures and water blooms of cyanobacteria using Microcystis aeruginosa peptide toxin polyclonal antibodies. Environ. Technol. 1988;9:1343–1348. doi: 10.1080/09593338809384699. [DOI] [Google Scholar]
- 19.Aranda-Rodriguez R., Jin Z., Harvie J., Cabecinha A. Evaluation of three field test kits to detect microcystins from a public health perspective. Harmful Algae. 2015;42:34–42. doi: 10.1016/j.hal.2015.01.001. [DOI] [Google Scholar]
- 20.Humpage A.R., Froscio S.M., Lau H., Murphy D., Blackbeard J. Evaluation of the abraxis strip test for Microcystins™ for use with wastewater effluent and reservoir water. Water Res. 2012;46:1556–1565. doi: 10.1016/j.watres.2011.12.015. [DOI] [PubMed] [Google Scholar]
- 21.Posthuma-Trumpie G.A., Korf J., van Amerongen A. Lateral flow (immuno) assay: Its strengths, weaknesses, opportunities and threats. A literature survey. Anal. Bioanal. Chem. 2009;393:569–582. doi: 10.1007/s00216-008-2287-2. [DOI] [PubMed] [Google Scholar]
- 22.Xing C., Liu L., Song S., Feng M., Kuang H., Xu C. Ultrasensitive immunochromatographic assay for the simultaneous detection of five chemicals in drinking water. Biosens. Bioelectron. 2015;66:445–453. doi: 10.1016/j.bios.2014.12.004. [DOI] [PubMed] [Google Scholar]
- 23.Tippkötter N., Stückmann H., Kroll S., Winkelmann G., Noack U., Scheper T., Ulber R. A semi-quantitative dipstick assay for microcystin. Anal. Bioanal. Chem. 2009;394:863–869. doi: 10.1007/s00216-009-2750-8. [DOI] [PubMed] [Google Scholar]
- 24.Spoof L., Meriluoto J. Cyanobacterial Monitoring and Cyanotoxin Analysis. John Wiley & Sons; Hoboken, NJ, USA: 2005. List of standard operating procedures (SOPs). Toxic; pp. 59–149. [Google Scholar]
- 25.Akter S., Vehniäinen M., Spoof L., Nybom S., Meriluoto J., Lamminmäki U. Broad-spectrum noncompetitive immunocomplex immunoassay for cyanobacterial peptide hepatotoxins (microcystins and nodularins) Anal. Chem. 2016;88:10080–10087. doi: 10.1021/acs.analchem.6b02470. [DOI] [PubMed] [Google Scholar]
- 26.Akter S., Vehniäinen M., Meriluoto J., Spoof L., Lamminmäki U. Non-competitive ELISA with broad specificity for microcystins and nodularins. Adv. Oceanogr. Limnol. 2017;8:121–130. doi: 10.4081/aiol.2017.6349. [DOI] [Google Scholar]
- 27.Hautala H., Lamminmäki U., Spoof L., Nybom S., Meriluoto J., Vehniäinen M. Quantitative PCR detection and improved sample preparation of microcystin-producing anabaena, microcystis and planktothrix. Ecotoxicol. Environ. Saf. 2013;87:49–56. doi: 10.1016/j.ecoenv.2012.10.008. [DOI] [PubMed] [Google Scholar]
- 28.Panferov V.G., Safenkova I.V., Varitsev Y.A., Zherdev A.V., Dzantiev B.B. Enhancement of lateral flow immunoassay by alkaline phosphatase: A simple and highly sensitive test for potato virus X. Microchim. Acta. 2018;185:25. doi: 10.1007/s00604-017-2595-3. [DOI] [PubMed] [Google Scholar]
- 29.Tang Z., Wang X., Lv J., Hu X., Liu X. One-step detection of ochratoxin A in cereal by dot immunoassay using a nanobody-alkaline phosphatase fusion protein. Food Control. 2018;92:430–436. doi: 10.1016/j.foodcont.2018.05.013. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.