Table 1. Operation Parameters of Intrepid XSP Radial ICP–OES.
| operating conditions | |
|---|---|
| incident power (W) | 1150 W |
| frequency of rf generator | 27.12 MHz |
| outer argon flow rate (L min–1) | 14 |
| auxiliary gas flow rate (L min–1) | 0.5 |
| carrier gas flow rate (L min–1) | 0.5 |
| solution uptake rate (L min–1) | 1 |
| Cu 324.754 | |
| analytical wavelength (nm) | Pb 220.353 |