Skip to main content
. Author manuscript; available in PMC: 2019 Jul 23.
Published in final edited form as: Clin Chem. 2019 Jan 15;65(3):406–418. doi: 10.1373/clinchem.2018.288092

Table 1.

Interassay imprecision forserum QCs.

Validation (n = 120)
Multiinstrument
validationa (n = 75)
Analyte Level Mean concentration, μg/mL CV, % Mean concentration, μg/mL CV, %
Acetylthreonine Low       0.0774 5.2       0.0775 5.8
Mid       0.892 5.1       0.867 5.5
High    1.78 4.8    1.69 4.8
Phenylacetylglutamine Low    0.329 5.2    0.337 4.2
Mid     8.17 4.0     8.30 4.5
High 16.0 3.8 16.1 4.8
Pseudouridine Low     1.27 5.2     1.30 6.9
Mid 16.4 4.2 16.5 5.7
High 32.1 3.7 31.7 5.6
Tryptophan Low     4.08 6.4     4.13 4.6
Mid 44.7 4.3 44.6 4.0
High 88.0 4.3 87.3 4.7
a

Data reported are from all 3 LC-MS/MSsystemstested.