Skip to main content
. 2011 Sep 14;25(5):324–329. doi: 10.1002/jcla.20477

Table 3.

Interference Test of Four Methods

Hemoglobin concentration (g/l) Nephelometry (%) Turbidimetry (%) Immuno‐colloidal gold method (%) RIA (%)
0.11  −6.9 −1.8 −10.1 −2.2
0.23 −7.7 −2.6 −10.1 −4.3
0.46 −7.7 1.7 −11.9 −10.0
0.61 −8.4 6.7 −17.4 −8.7
0.91 −8.8 14.1 −13.8 −10.4
1.82 −9.5 26.3 −17.8 −11.8
2.74 −11.4 27.8 −21.1 −12.5