Skip to main content
. 2019 Jul 17;13:24. doi: 10.3389/fnint.2019.00024

TABLE 7.

Inter and intra-scanner variability of MD in white matter ROIs.

Label Measure Mean MD SD MD CV (%) MD Inter: intra CV ratio Mean MD SD MD CV (%) MD Inter: Intra CV ratio
LEFT RIGHT
Anterior limb internal capsule Inter-scanner 7.3 0.6 8.2 5.9 7.4 0.6 8.1 5.8
Mean intra-scanner 7.2 0.1 1.4 7.3 0.1 1.4
Arcuate fasciculus region 1 Inter-scanner 7.2 0.4 5.6 4.0 7.3 0.4 5.5 3.9
Mean intra-scanner 7.2 0.1 1.4 7.3 0.1 1.4
Arcuate fasciculus region 2 Inter-scanner 7.5 0.3 4.0 3.1 7.3 0.2 2.7 1.9
Mean intra-scanner 7.5 0.1 1.3 7.3 0.1 1.4
Arcuate fasciculus region 3 Inter-scanner 7.5 0.3 4.0 3.1 7.6 0.3 3.9 3.0
Mean intra-scanner 7.5 0.1 1.3 7.6 0.1 1.3
Cingulum Inter-scanner 7.6 0.4 5.3 4.1 7.5 0.4 5.3 4.1
Mean intra-scanner 7.6 0.1 1.3 7.5 0.1 1.3
Posterior limb internal capsule Inter-scanner 7.2 0.5 6.9 4.9 7.0 0.4 5.7 4.1
Mean intra-scanner 7.2 0.1 1.4 7.0 0.1 1.4
Sagittal stratum Inter-scanner 8.3 0.5 6.0 2.5 8.1 0.4 4.9 4.1
Mean intra-scanner 8.3 0.2 2.4 8.1 0.1 1.2
Uncinate fasciculus Inter-scanner 8.0 0.4 5.0 2.0 8.1 0.3 3.7 1.5
Mean intra-scanner 8.1 0.2 2.5 8.1 0.2 2.5

BILATERAL
Corpus callosum Inter-scanner 9.0 0.6 6.7 6.1
Mean intra-scanner 9.0 0.1 1.1

All scans were included (n = 24). DTI data were not available Scanner B. MD is scaled × 10,000 mm2/s.