Skip to main content
. 2019 Apr 24;170(2):345–356. doi: 10.1093/toxsci/kfz102

Table 2.

Tolerance Intervals (TI 90%–95%) for CTD90, Fridericia’s Rate-Corrected CTD90 (CTD90c F), Beat Rate, and Amplitude of the Calcium Transient in hSC-CMs at Each Test Site

Parameter Site
Ncardia
Janssen
MSD
Lower Upper Lower Upper Lower Upper
CTD90 −5 5 −4 4 −10 10
CTD90c F −4 4 −3 3 −8 8
Beat rate −8 8 −7 7 −16 16
Amplitude −5 5 −7 7 −14 14