Skip to main content
. 2019 Feb 28;12(2):367–379. doi: 10.15283/ijsc18151

Table 2.

Troubleshooting table

Step Problem Possible reason Solution
2,4,5 Cell death occurs High concentration of FreSHtracer Test the toxicity of FreSHtracer in cells of interest and load probes in the non-toxic concentration range
Phototoxicity Adjust laser power or time interval for image acquisition
2-B-i MitoFreSHtracer stains mitochondria unevenly Low solubility of MitoFreSHtracer in medium Add 0.03% Pluronic F-127 to the culture medium
4,5,13 Fluorescence intensity is low Background fluorescence of the culture medium Use Phenol red-free medium or buffer solutions during probe loading and observation
4,5 Basal fluorescence intensity is not constant Laser power unstable Wait 20~60 min before image acquisition
Temperature unstable Incubate cells at 37°C during the entire experimental procedure
10,13,27 FR measurements are not consistent Cellular GSH concentration is affected by thermal stress Maintain cells at a constant temperature (4°C) until flow cytometer analysis